Download Print this page
Mastercare Gen II Entree Bath Maintenance Manual

Mastercare Gen II Entree Bath Maintenance Manual

Hide thumbs Also See for Gen II Entree Bath:

Advertisement

Quick Links

Gen II Entree Bath
All Models
Maintenance Manual
Important Safety Instructions
Read & Follow All Instructions Thoroughly
Note: The serial tag is at the first page of this manual package. The Standards and
certification labels are located in the Architectural Specifications & Rough-In
Information

Advertisement

loading
Need help?

Need help?

Do you have a question about the Gen II Entree Bath and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Mastercare Gen II Entree Bath

  • Page 1 Gen II Entree Bath All Models Maintenance Manual Important Safety Instructions Read & Follow All Instructions Thoroughly Note: The serial tag is at the first page of this manual package. The Standards and certification labels are located in the Architectural Specifications & Rough-In...
  • Page 3 Generation II Entree Bath – All Models MAINTENANCE MANUAL MasterCare hygiene products are designed and manufactured to help provide a more operationally efficient and secure environment for your staff and residents. It is required that a scheduled maintenance program be set up to insure that the efficiencies and security designed into the bathing system are kept at peak levels.
  • Page 4 Generation II Entree Bath – All Models MAINTENANCE MANUAL PROBLEM SOLVING AND DIAGNOSIS: The following guide is intended to help solve the majority of problems that may occur due to normal wear and use of the system over time. Problem Potential Causes Door does not open/close Procedure: Drain must be fully open for door to open.
  • Page 5 The door is out of adjustment or the seal is flattened and unable to compress further. To correct: call MasterCare to order a new seal and replacement instructions. 5. Seal is worn out or damaged: The inner core of the seal is permanently inflated with a material called closed cell foam and springs back to a round shape after the door opens.
  • Page 6 Generation II Entree Bath – All Models MAINTENANCE MANUAL DOOR ADJUSTMENT DIAGRAM Actuator Motor Door open/close adjustment screw. Using a 1/8 x 6” flat head screwdriver rotate clockwise to open more. Make adjustments in 1/32 turns. Do not over-rotate the door CAUTION: open position.
  • Page 7 Generation II Entree Bath – All Models MAINTENANCE MANUAL ACTUATOR - GEN II NON-ELECTRIC DIAGRAM 2071 14 Ave – Columbus, NE 68601 07-01-16 800-798-5867 FAX: 402-563-9102 Page 7 of 17 www.mastercarebath.com...
  • Page 8 Generation II Entree Bath – All Models MAINTENANCE MANUAL ELECTRICAL ASSEMBLY DIAGRAM - 1 2071 14 Ave – Columbus, NE 68601 07-01-16 800-798-5867 FAX: 402-563-9102 Page 8 of 17 www.mastercarebath.com...
  • Page 9 Generation II Entree Bath – All Models MAINTENANCE MANUAL ELECTRICAL ASSEMBLY DIAGRAM - 2 2071 14 Ave – Columbus, NE 68601 07-01-16 800-798-5867 FAX: 402-563-9102 Page 9 of 17 www.mastercarebath.com...
  • Page 10 Generation II Entree Bath – All Models MAINTENANCE MANUAL ENTREE DOOR SEAL MAINTENANCE & REPLACEMENT INSTRUCTIONS (Parts #211081 & 211082): Note: Read all instructions carefully and thoroughly. Refer to the attached Entree Door Seal drawing to reference details. 6. Push door close button, bringing door up and stop once the Phillips head screw are visible on the side of the door.
  • Page 11 Generation II Entree Bath – All Models MAINTENANCE MANUAL ENTREE DOOR SEAL DIAGRAM 2071 14 Ave – Columbus, NE 68601 07-01-16 800-798-5867 FAX: 402-563-9102 Page 11 of 17 www.mastercarebath.com...
  • Page 12 Generation II Entree Bath – All Models MAINTENANCE MANUAL AERATION INJECTOR ASSY DIAGRAM 2071 14 Ave – Columbus, NE 68601 07-01-16 800-798-5867 FAX: 402-563-9102 Page 12 of 17 www.mastercarebath.com...
  • Page 13 Generation II Entree Bath – All Models MAINTENANCE MANUAL DRAIN SHOE-ENTREE REPLACEMENT DIAGRAM 2071 14 Ave – Columbus, NE 68601 07-01-16 800-798-5867 FAX: 402-563-9102 Page 13 of 17 www.mastercarebath.com...
  • Page 14 Generation II Entree Bath – All Models MAINTENANCE MANUAL GEN II MIXING VALVE CARTRIDGE Adjustment & Replacement Instructions (Item #600578): Note: Read and follow all instructions thoroughly. Refer to the GEN II Mixing Valve Diagram to reference details. Supplies needed: Silicone lubricant if replacing the cartridge (readily available at local hardware stores). Adjustment Instructions: 20.
  • Page 15 Generation II Entree Bath – All Models MAINTENANCE MANUAL GEN II MIXING VALVE DIAGRAM 2071 14 Ave – Columbus, NE 68601 07-01-16 800-798-5867 FAX: 402-563-9102 Page 15 of 17 www.mastercarebath.com...
  • Page 16 Generation II Entree Bath – All Models MAINTENANCE MANUAL DOOR OPEN/CLOSE BUTTONS INSTRUCTIONS (Part #706127): Note: Read and follow all instructions thoroughly. Refer to the drawing bellow to reference. 1. Loosen and remove both thumb screws from the foot end control. Gently tilt the foot end control to the side. 2.
  • Page 17 Generation II Entree Bath – All Models Contact Contact FOR ADDITIONAL MAINTENANCE HELP OR PLACE AN ORDER: Online: www.mastercarebath.com, Email: mastercareweb@mastercarebath.com Call: 1-800-798-5867 (M-F/8a.m .- 5p.m. CST) 2071 14 Ave – Columbus, NE 68601 07-01-16 800-798-5867 FAX: 402-563-9102 Page 17 of 17 www.mastercarebath.com...