Auto (Auto Settings) Strategy - UNI-T UTD2000M Operating Manual

Digital storage oscilloscope
Table of Contents

Advertisement

 
UTD2000M User Manual     
Function 
Bandwidth Limit   
Vertical Scale Coefficient   
Volts/Grid 
Phase Reversed 
Horizontal Position 
Second/Grid 
Acquiring Mode 
Sampling Mode     
Fast Sampling 
Trigger Type 
Trigger Signal Source     
Trigger Coupling 
Trigger Mode 
Hold Off   
Trigger Level 
Slope Type 
Signal Identification     
 
Remarks: 
The adjustment of trigger signal source while the Unit is in "AUTO" is based on the following points:         
        1. If only one channel has input signal, then set the signal input channel as the trigger source.     
        2. When more than one channel are used, the auto setting function sets the vertical control of each channel 
and uses the active channel with minimal number to set horizontal and trigger control.         
        3. If no signal has been found in all channels, then use CH1 channel as the trigger source.     
2.11.2    Auto (Auto Settings) Strategy 
In actual use, many times, we hope the unit is able to auto set itself the way we want it be. In this sense, auto (auto 
 
 
 
Open 
Full bandwidth 
Adjust in accordance with signal 
amplitude     
Rough tuning   
Close     
Auto Adjustment 
Adjust in accordance with signal 
frequency 
Normal sampling   
Equivalent or maintain current 
settings real‐timely   
Close 
Edge 
According to signal of CH1 or CH2, see 
remarks       
Maintain current setting while the 
unit is in DC, AC, Low frequency reject 
or high frequency reject mode;   
Auto   
Minimum Value 
Set to be 50% 
Rising 
Automatically adjust the channel 
60
Lock     
Full bandwidth or maintain current 
settings of 20MHz.   
Adjust in accordance with signal 
amplitude     
Rough tuning 
Open, close or maintain current settings 
Auto Adjustment 
Adjust in accordance with signal 
frequency 
Maintain one current setting while the 
unit is in normal sampling, peak value or   
average state; 
Equivalent or maintain current settings 
real‐timely 
Close 
Maintain current setting while the unit 
is in edge, pulse width, video or slope 
triggering mode.           
Maintain current setting while the unit 
is in CH1, CH2, EXT, EXT/5 or AC Line 
mode.   
Maintain current setting while the unit 
is in DC, AC, Low frequency reject or 
high frequency reject mode; 
Auto 
Maintain current setting 
Set to be 50% 
Maintain current setting 
Currently opened channel 

Advertisement

Table of Contents
loading

Table of Contents