UTD2000M User Manual     
to acquire signal baseline so as to confirm whether the acquisition is normal;                   
After pressing SINGLE key to set single trigger, FORCE (force trigger) key can be pressed to execute 
experimental acquisition so as to check the control settings.             
Trigger Control MENU   
Trigger Type: Edge, Pulse width, video, slope, less amplitude pulse, alternative             
Edge Triggering: Trigger occurs when the edge of a triggering signal reaches a given level.   
Pulse Width Triggering: Trigger occurs when the pulse width of a triggering signal reaches a preset triggering 
condition.                   
Video Triggering: Perform field triggering or row triggering to standard video signal.     
Slope Triggering: Trigger occurs when the slope of triggering signal meets the condition.               
Less Amplitude Pulse Triggering: Trigger the pulse signal that has passed one trigger level but has failed to pass 
another one.           
Alternative Triggering: Trigger signals that has no frequency connections.         
2.4.1      Edge Triggering 
Edge triggering refers to a trigger that occurs at the edge of an input signal, namely the trigger threshold value. 
When "edge triggering" is selected, it means the trigger occurs at the rising edge and falling edge of an input 
signal.             
 
Table 2‐8    Edge Triggering 
Menu 
Type 
Signal Source 
Trigger Coupling   
Triggering Mode 
 
 
 
Setting 
Edge 
CH1、CH2 
EXT、EXT/5 
AC LINE 
Alter 
DC 
AC 
Low Frequency Reject   
High Frequency Reject 
Auto 
 
Normal 
 
Description   
 
Set CH1 or CH2 as the source of trigging signal 
Set external triggering or the quotient of dividing external 
triggering signal by 5 as the signal source           
Set mains supply as the source of trigger 
Alternatively use CH1 or CH2 as the source of trigger     
Pass both the AC and DC component in an input signal;   
Block the DC component in an input signal;           
Reject the low frequency component (less than 80KHz) contained 
in a signal;       
Reject the high frequency component (higher than 80KHz) 
contained in a signal; 
When there is no input of trigger signal, the system will 
automatically acquire waveform data and display scanning 
baseline on the screen; when there is a trigger signal, it will then 
automatically switch into trigger scanning.           