Gima 28209 Manual

Gima 28209 Manual

Vega uno suction aspirator
Hide thumbs Also See for 28209:

Advertisement

Quick Links

PROFESSIONAL MEDICAL PRODUCTS
ASPIRATORE CHIRURGICO VEGA UNO
VEGA UNO SUCTION ASPIRATOR
ASPIRATEUR VEGA UNO
CHIRURGISCHER ABSAUGER VEGA UNO
ASPIRADOR QUIRÚRGICO VEGA UNO
APARELHO DE SUCḈÃO CIRÚRGICO VEGA UNO
28209
Gima S.p.A.
Via Marconi, 1 - 20060 Gessate (MI) Italy
gima@gimaitaly.com - export@gimaitaly.com
www.gimaitaly.com
Made in Italy
0476

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 28209 and is the answer not in the manual?

Questions and answers

Summary of Contents for Gima 28209

  • Page 1 VEGA UNO SUCTION ASPIRATOR ASPIRATEUR VEGA UNO CHIRURGISCHER ABSAUGER VEGA UNO ASPIRADOR QUIRÚRGICO VEGA UNO APARELHO DE SUCḈÃO CIRÚRGICO VEGA UNO 28209 0476 Gima S.p.A. Via Marconi, 1 - 20060 Gessate (MI) Italy gima@gimaitaly.com - export@gimaitaly.com www.gimaitaly.com Made in Italy...
  • Page 2: Important Safety Rules

    GIMA technical service department. Do not attempt to make the device work before it has been thorough- ly checked by qualified personnel and/or the GIMA technical service department.
  • Page 3: Technical Characteristics

    7. Instrument and accessory discharging must be done according to current regulations in the coun- try of use. 8. WARNING: Do not change this equipment without the permission of the manufacturer GIMA S.p.A. None of electric or mechanical parts has been designed to be repaired by customers or end-users. Don’t open the device, do not mishandle the electric / mechanical parts.
  • Page 4: Cleaning Of The Device

    ENGLISH Working Condition Room temperature: 5 ÷ 35°C Room humidity percentage: 10 ÷ 93% RH Atmospheric pressure: 700 ÷ 1060 hPa Conservation condition and Transport Room temperature: -25 ÷ 70°C Room humidity percentage: 0 ÷ 93% RH Atmospheric pressure: 500 ÷ 1060 hPa The technical specifications may change without notice.
  • Page 5: Cleaning Of Accessories

    ENGLISH whether the fitting is suitable for reuse. The component must be replaced if there are visible signs of decay in the material constituting the said component. Service life of the device: More than 850 hours of operation (or 3 years) in accordance with the standard conditions of testing and operation.
  • Page 6 2° case – If both the security system doesn’t work, there is the possibility that liquid comes inside the device, in this case return the device to GIMA technical service. Gima S.p.A. will provide upon request electric diagrams, components list, descriptions, setting in- structions and any other information that can help the technical assistance staff for product repair.
  • Page 7 ENGLISH INSTRUCTION FOR USE • The device must be checked before each use in order to detect malfunctions and / or damage caused by transport and / or storage. • The working position must be such as to allow one to reach the control panel and to have a good view of the empty indicator, the jar and the antibacterial filter.
  • Page 8 ENGLISH Antibacterial Filter VACUUM Port PATIENT Port Conical Fittings Autoclavable Silicone Tube Vacuum Vacuum Indicator Regulator Knob (kPa /Bar) ON/OFF Switch NEVER USE THE DEVICE WITHOUT JAR AND / OR PROTECTION FILTER MAKE SURE THAT CHILDREN AND/OR MENTALLY ILL PEOPLE DO NOT USE THE DEVICE WITHOUT ADULT SURVEILLANCE RISK OF ELECTROMAGNETIC INTERFERENCE AND POSSIBLE REMEDIES This section contains information regarding the conformity of the compliance with the EN 60601-1-2 Standard...
  • Page 9 ENGLISH Guidance and manufacturer’s declaration – Immunity Emissions The VEGA UNO SUCTION UNIT is intended for use in the electromagnetic environment specified below. The customers or the user of the VEGA UNO SUCTION UNIT should make sure that it’s used in such an environment. Immunity Test Level indicated by the Compliance Level...
  • Page 10 ENGLISH Where P is the maximum nominal output voltage of the transmitter in Watt (W) depending on the manufactur- er of the transmitter and the recommended separation distance in metres (m). The intensity of the field from the fixed RF transmitters, as determined by an elec- tro-magnetic study of the site , could be lower than the level of conformity of each frequency interval...
  • Page 11 Disposal: The product must not be disposed of along with other domestic waste. The users must dispose of this equipment by bringing it to a specific recycling point for electric and electronic equipment. GIMA WARRANTY TERMS The Gima 12-month standard B2B warranty applies.