Xylem WTW TriOxmatic 690 Operating Manual page 27

Dissolved oxygen sensor
Table of Contents

Advertisement

®
TriOxmatic
690
Variant 2:
Cleaning the silver
counter electrode in a
beaker
ba25314e05
09/2020
1
Wipe the silver counter electrode with a lint free paper towel
and carefully remove any loose deposits.
2
Fill a beaker (150 ml, high-sided form) with approx. 25 ml RL-
AG/Oxi cleaning solution.
NOTE
The reference electrode must not come into contact with the cleaning
solution under any circumstances. This could destroy the reference
electrode and cause the sensor to become defective.
3
Clamp the sensor in a stand.
Submerse the electrode unit in the RL-AG/Oxi cleaning
solution to just above the silver counter electrode.