SEMI-AUTOMATED URINE ANALYZER AUTION ELEVEN ™ AE-4022 Table of Contents This Quick Reference Guide is a brief System Overview ......1 summary of the operation and routine Sample Preparation ....2 maintenance of the AUTION ELEVEN Sample Measurement ....3-5 Semi-Automated Urine Analyzer.
Page 4
SEMI-AUTOMATED URINE ANALYZER AUTION ELEVEN System Overview System Overview ™ AE-4022 Name of each part Setup Built-in printer Select t Display Operator panel Maintenance cover Set the Test strip tray Waste box Carrying arm Cover open button Set the (on both sides) Test strip detection window Meas...
Page 6
Test strip tray Waste box Carrying arm Cover open button (on both sides) Test strip detection window SEMI-AUTOMATED URINE ANALYZER AUTION ELEVEN Sample Preparation ™ AE-4022 Daily Maintenance Turn off the instrument. Open the maintenanc Rails (white) 1. Turn on power switch at back of instrument. Measurement Preparation for measurement Preparation of measurement...
Page 8
Call up the screen a With the [Standby screen] displayed, press and hold for 3 Chapter 2 Measurement Operation seconds. • The [Operator ID number input screen] appears. 2.3.6 Entering patient ID numbers SEMI-AUTOMATED URINE ANALYZER AUTION ELEVEN Sample Measurement ™...
Page 9
Close the test strip bottle cap securely. IMPORTANT: After taking the test strips out ement Operation of the bottle, immediately close the cap or the test strips in the bottle will absorb moisture or dirt Take out the required number of test strips from in the air, which may deteriorate Measurement operation the bottle.
Page 10
Measurement Dip the entire test strip pad area in the sample for two seconds. IMPORTANT Do NOT wet the Do NOT wet the black mark. black mark. After dipping the test strip in the sample, SEMI-AUTOMATED URINE ANALYZER AUTION ELEVEN Sample Measurement ™...
Page 12
Before using a Control, read carefully its package insert. of the bottle, immediately close the cap or the test strips in the a Prepare the Control. bottle will absorb moisture or dirt in the air, which may deteriorate reagents and make the test strips useless.
Page 14
Do NOT wet the a Place a white check strip on the test strip tray. • If any of the measurement results are still outside the ranges, contact ARKRAY Technical Support at 1.855.646.3108. Do NOT wet the black mark. Do NOT wet the black mark.
Lever Daily Maintenance Detach the test strip tray. Power switch Turn off the instrument. Open the maintenance cover. Detach the fe Test strip tray Feed lever in the center SEMI-AUTOMATED URINE ANALYZER AUTION ELEVEN Maintenance – Daily Discard used test strips. ™...
Page 17
Carrying arm test strip tray. trip tray Feed lever Feed lever SEMI-AUTOMATED URINE ANALYZER AUTION ELEVEN Maintenance – Daily ™ AE-4022 Clean the Feeder Wipe the parts attaching them 8. Clean the table. 7. Detach feed lever by sliding • Clean the areas circled in red below with water. the white lever in the center suction port.
Page 18
Open the maintenance cover. Detach the feed lever by sliding the white lever Wash the detached carrying arm, waste in the center towards you. test strip tray, suction port, test strip feed SEMI-AUTOMATED URINE ANALYZER AUTION ELEVEN Maintenance – Daily and feed lever with water to remove any ™...
asurements. Discard the used test strips, and sterilize and clean the Items required: Alcohol, cloth, and protective gloves. exposure to pathogenic microorganisms. Discard used test strips ance with local regulations for biohazardous waste. a Check that the [Standby screen] is displayed. SEMI-AUTOMATED URINE ANALYZER AUTION ELEVEN Maintenance –...
Cut the thermal recording paper from the printer, the screen message shown in the right figure appears and a continuous beep sounds for Cut the thermal recording paper a Check that the [Standby screen] is displayed. approximately 1 minute. The alarm sound can be stopped c If some thermal recording paper still remains in the printer, cut Take out.
Page 22
W005 Wrong test strip A test strip other than that specified Only use ARKRA Only use ARKRAY’s test strips designed for was used. designed for the W005 Wrong test strip A test strip other than that specifi ed was used.
Page 24
To prevent the error from recurring, take appropriate measures (corrective action), referring to the table below. If unable to clear the error, note error code and call ARKRAY Technical Support at 1.855.646.3108. Press to cancel the error notice. The display returns to the [Standby screen].
Page 25
SEMI-AUTOMATED URINE ANALYZER AUTION ELEVEN Errors ™ AE-4022 Message Error Possible cause Measures (Corrective Action) E007 Data not No data was found in the specifi ed range. Verify that the specifi ed range for re-printing, found re-sending, or list printing is appropriate. E008 Auto start •...