The following optional components are compatible with Stellar:
• 3 m air tubing • SlimLine™ air tubing • Clear air tubing (disposable) • H4i™ heated
humidifier • Antibacterial filter • Heat moisture exchanger filter (HMEF) • ResMed XPOD
oximeter • Nonin™ pulse oximetry sensors • FiO
adapter) • FiO
WARNING
The Stellar should only be used with air tubing and accessories recommended by
ResMed. Connection of other air tubing or accessories could result in injury or
damage to the device.
ResMed regularly releases new products. Please check the catalogue of ventilation
accessories on our website at www.resmed.com.

Patient interface

Both masks and tracheostomy tubes can be used with Stellar. To set the patient interface
type go to the Setup menu, select Clinical Settings, then Advanced Settings.
For information on using masks, see the mask manual. For a full list of compatible masks
for this device, see the Mask/Device Compatibility List on www.resmed.com on the
Products page under Service & Support. If you do not have internet access, please
contact your ResMed representative.

Humidification

A humidifier is recommended especially for patients experiencing dryness of the nose,
throat or mouth. For information on using a humidifier as part of:
noninvasive ventilation, see "Setting up for noninvasive use" on page 3.
invasive ventilation, see "Setting up for invasive use" on page 4.

Internal battery

CAUTION
Over time, the capacity of the internal battery will decrease. This depends on
individual use and ambient conditions. As the battery degrades, the battery charge
level indicator may lose accuracy. This will mean battery-related and system alarms
may not activate as intended. ResMed recommends the battery be tested after two
years to assess the remaining battery life. Testing and replacement of the internal
battery should only be performed by an authorised service agent.
Note: The battery duration depends on the state of charge, the environmental conditions,
the condition and age of battery, the device settings and the patient circuit configuration.
In case of a mains power disruption, the device will operate using the internal battery if
there is no external battery connected to the device. The internal battery will operate for
approximately two hours under normal conditions (see "Technical specifications" on
page 25). The power status of the battery is displayed on top of the LCD screen. Check the
battery status regularly while operating the device with the internal battery and connect the
device in time to mains power or alternatively to the external battery.
Additionally the Internal battery use alarm will be displayed. Press the Alarm mute button
to clear the alarm.
To recharge the internal battery, connect the device to mains power. It can take up to three
hours to fully recharge the internal battery, however this can vary depending on
environmental conditions and if the device is in use.
Maintenance
The internal battery has to be discharged and recharged every six months.
1
Remove the power cord while the Stellar is turned on (stand-by or in operation) and let the
device operate with the internal battery to a charge level of 50%.
2
Reconnect the power cord to the mains power. The internal battery will be fully recharged.
2
monitoring sensor • Stellar Mobility bag • ResMed Leak Valve • Tubing wrap.