Set Mains Voltage - Baur PGK 50 Operating Instructions Manual

High voltage testing sets
Table of Contents

Advertisement

5. Servicing

Set mains voltage

Selectable mains voltages
I
100µA
1mA
10µA
10mA
1µA
F1
F2
PGK
F1
F2
Component layout of power PCB
Component layout of voltage supply PCB
Mains voltage
110 V
115 V
120 V
127 V
220 V
230 V
240 V
5-2
The PGK50/80 mains voltage is factory installed according to
customer requirements. Jumper locations and fuses for the
appropriate mains voltages are provided as follows:
U [k
8 10 12 14 16
6
18
4
20
2
100mA
22
0
24
26
28
30
T / min
Position
Position
plug X15
plug X1
110 V
110 / 115 V
110 V
110 / 115 V
110 V
120 / 127 V
110 V
120 / 127 V
220 V
220 / 230 V
220 V
220 / 230 V
220 V
240 V
Check set mains voltage on type plate and make corrections if
required.
X15
Mains fuse
F1, F2 (ø6.3x32)
8 AT
8 AT
8 AT
8 AT
6.3 AT
6.3 AT
6.3 AT
F3
X1
F4
Fuse
Fuse
F4 (ø5x20)
F3 (ø5x20)
0.8 AT
6.3 A fast
0.8 AT
6.3 A fast
0.8 AT
6.3 A fast
0.8 AT
6.3 A fast
0.4 AT
6.3 A fast
0.4 AT
6.3 A fast
0.4 AT
6.3 A fast