b. Disconnect the output of the BNC termination from the
VERT INPUT connector and reconnect it to the EXT TRIG OR
HORIZ INPUT connector.
c. Set the square-wave generator for a five-division dis­
play on the test oscilloscope at one kilohertz.
d. CHECK—Test oscilloscope display for optimum square-
wave response (optimum square corner and flat top).
e. ADJUST— C206 (see Fig. 6-11C) for optimum square-
wave response.
f. Disconnect all test equipment.

27. Adjust High-Frequency Compensation

a. Change the following control settings:
VOLTS/DIV
Trig/Horiz Coupling
TIME/DIV
X 10 HORIZ MAG
b.
Connect the square-wave generator fast-rise +
put connector to the VERT INPUT connector through the GR
to BNC adapter, 42-inch 50-ohm BNC cable, 10X BNC
attenuator and the 50-ohm BNC termination.
c. Set the square-wave generator for a four-division dis­
play at 100 kilohertz.
d. Move the leading edge of the square wave onto the
viewing area with the horizontal POSITION control.
e. Set the test oscilloscope for a vertical deflection fac­
tor of one volt/division (10 volts/division at 10X probe tip)
at a sweep rate of two microseconds/division. Adjust the
triggering controls when necessary to provide a stable dis­
play.
f. Connect the 10X probe tip to the collector of Q163
(case of Q163; see Fig. 6-12B). Be sure the probe is com­
pensated.
g. CHECK—Test oscilloscope display for flat top on
square wave similar to Fig. 6-12A.
h. ADJUST—Cl 60 (see Fig. 6-12B) for optimum flat top
on square wave.
i. Connect the 10X probe tip to the collector of Q173
(case of Q173; see Fig. 6-12B).
j. CHECK—Test oscilloscope display for flat bottom on
square wave similar to Fig. 6-12A.
k. ADJUST—C l70 (see Fig. 6-12B) for optimum flat bottom
on square wave.
. Install the calibration shield on the Type 323.
l
m. CHECK—Type 323 CRT display for optimum square
corner similar to Fig. 6-12C.
o
.01
INT TRIG AC
5 /xs
Pulled out
out­
n. ADJUST— Cl 60 and Cl 70 about the same amount in
the same direction for optimum square corner on the CRT
display.
o. Disconnect all test equipment and remove the cali­
bration shield.
Fig. 6 -1 2 .
(A )
Typical oscilloscope d ispla y when adjusting
frequency com pensation, (B) location o f high-frequency compensation
test points and adjustments (M a in b o a rd ), (C) typica l CRT display
showing correct high-frequency com pensation.
Calibration— Type 323
high-
6-17