Jumper Settings

On the Remote Indicator there is a failsafe jumper and a write protect jumper behind the
display on the Communication Module. The top jumper is the failsafe jumper. It is highly
recommended to put the failsafe jumper to DOWN (Downscale). The bottom jumper sets
the write protect.
ATTENTION: Electrostatic Discharge (ESD) hazards. Observe
precautions for handling electrostatic sensitive devices.
Step
Action
Ensure the Enraf FlexLine Gauge is switched off or that the Remote
1
Indicator is disconnected.
Loosen the end-cap lock, and unscrew the end cap from the
2
Electronics side of the Transmitter housing.
Carefully depress the tabs on the sides of the Display Module and pull
3
it off.
Set the write protect jumper and the failsafe jumper to the desired
4
behavior. See Table 1 for jumper positioning.
Reinstall the Display Module. Carefully line up the display and
5
interface connector and snap it into place. Verify that the two tabs on
the sides of the display latch.
Screw on the end cap and tighten the end-cap lock.
6
Reconnect the Remote Indicator or switch on the Enraf FlexLine
7
Gauge.
Image
Description
Failsafe = DOWN (3.8mA)
Write Protect = OFF (Not Protected)
Failsafe = UP (21.8mA) NOT RECOMMEND
Write Protect = OFF (Not Protected)
Failsafe = DOWN (3.8mA)
Write Protect = ON (Protected)
Failsafe = UP (21.8mA) NOT RECOMMEND
Write Protect = ON (Protected)
Table 1: Jumper settings
<Return> Return to the Level 1 menu
<Return>
HART
7 or 8
HART polling address
Address
Default:
7
Enter the time in minutes
Parameters
until the device
0 – 15
automatically goes online
Standby
Default:
once in standby.
Time
5
0 means the device does
NOT automatically go
online.
Table 4: Device Setup Menu
<Return> Return to the level 1 menu
<Return>
Display
The firmware version of the
Firmware Version
Display Module
<Return>
Comm
The firmware version of the
Module
Firmware Version
Communication Module
Table 5: Information Menu
<Return> Return to the Level 1 menu
<Return>
Critical
OK
Comm Module
FAULT
<Return>
OK
Supply Voltage
LOW
Non-
HIGH
Critical
Comm Module
OK
Temp
OVER TEMP
Display Setup
OK
NVM
CORRUPT
Table 6: Diagnostic Menu
All Diagnostics menu items are Read Only
Press ↵ to edit,
↑ or ↓ to select
number and ↵ to
enter.
Press ↵ to edit,
↑ or ↓ to select
number and ↵ to
enter.
Read Only
Read Only
FAULT: There is a problem with the
Communications Module.
LOW: Supply voltage is below the
low specification limit
HIGH: Supply voltage is above the
high specification limit.
OVER TEMP: Communications
Module temperature is greater than
85°C.
NVM CORRUPT: The Display setup
memory is corrupted.
RMA805 Enraf FlexLine Remote Indicator Quick Start Guide

Configuration Guide

The Remote Indicator 3-button interface provides user interface and operation capability
without opening the Remote Indicator.
The user must press ↵ button to call up the Main Menu. To exit the Main Menu and
return to the PV display screen, select <EXIT>.
Use the  button to scroll through the list of menu items. Press the ↵ button to select an
item for data entry or activation. When an item is selected for data entry or activation, the
cursor is positioned over the left-most digit to allow editing of the value. No action is
taken against a menu item until the ↵ button is pressed.
Level 1
Level 2
<Exit>
n/a
Critical
Diagnostics
Non-Critical
Display Setup
LCD Contrast
HART Setup
Device Setup
Parameters
Display
Information
Comm Module
Table 2: Main Menu Structure
<Return> Return to the Level 1 menu
<Return>
LCD
Set
0 – 9
Contrast
Contrast
Default: 5
Table 3: Display Setup Menu

Appendix A. PRODUCT CERTIFICATIONS

A1. Hazardous Locations Certifications
MSG
AGENCY TYPE OF PROTECTION
CODE
Flame-proof and Dust:
II 2 G Ex db IIC T6..T5 Gb
II 2 D Ex tb IIIC T 95°C Db
Intrinsically Safe:
II 1 G Ex ia IIC T4 Ga
C
ATEX
Non-Incendive
II 3 G Ex ec IIC T4 Gc
Enclosure: Type IP66/IP67
STANDARDS: EN 60079-0: 2018; EN 60079-1: 2014; EN 60079-11: 2012; EN
60079-31: 2014; EN 60079-7: 2015;
Flame-proof:
Ex db IIC T6..T5 Gb
Ex tb IIIC T 95°C Db
Intrinsically Safe:
Ex ia IIC T4 Ga
D
IECEx
Non-Incendive
Ex ec IIC T4 Gc
Enclosure: IP66/ IP67
STANDARDS: IEC 60079-0: 2017 IEC 60079-1: 2014;
IEC 60079-11: 2011; IEC 60079-7: 2017; IEC 60079-31: 2013
Level 3
n/a
For details see table 6
For details see table 5
For details see table 4
For details see table 3
Press ↵ to edit,
↑ or ↓ to select
Adjust the LCD contrast
number and ↵ to
level.
enter.
Electrical
Ambient Temperature
Parameters
T6: -20°C to 65°C T95°C,
Note 1
T5: -20°C to 85°C
Note 2
-20°C to 70°C
Note 1
-20°C to 85°C
ALL
ALL
T6: -20°C to 65°C T95°C,
Note 1
T5: -20°C to 85°C
-20°C to 70°C
Note 2
Note 1
-20°C to 85°C
ALL
ALL