Page 1
Valor™ 2000W Series Instruction Manual Serie Valor™ 2000W Manual de instrucciones Série Valor™ 2000W Manuel d’instructions Valor™ 2000W Serie Bedienungsanleitung Serie Valor™ 2000W Manuale di istruzioni GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Valor™ 2000W Series EN-1 1. INTRODUCTION This manual contains installation, operation and maintenance instructions for the Valor™ 2000W Series. Please read the manual completely before using the scale. 1.1 Safety Precautions Please follow these safety precautions: • Verify that the AC Adapter input voltage matches the local AC power supply. •...
EN-2 Valor™ 2000W Series 2.3 Selecting the Location Use the scale on a clean, firm and flat surface. Avoid locations with excessive air current, vibrations, heat sources, or rapid temperature changes. 2.4 Leveling the Equipment Adjust the leveling feet so the bubble is centered in the circle. Be sure the equipment is level each time its location is changed.
Charging the scale must be performed in a dry environment. CAUTION: Battery is to be replaced only by an authorized OHAUS service dealer. Risk of explosion can occur if the rechargeable battery is replaced with the wrong type or if it is not properly connected.
Page 5
EN-4 Valor™ 2000W Series 3. OPERATION – V22PW 3.1 Controls Figure 3-1. Front Control Panel Valor 2000W – V22PW TABLE 3-1 Button Functions Short Press (when on): Sets display to zero On/Zero Short Press (when off): Turns scale on Long Press (when on): Turns the scale off Short Press (in Menu): Selects / accepts displayed setting Tare...
Valor™ 2000W Series EN-5 TABLE 3-2 Display Symbols Item Description Item Description NET symbol Dynamic symbol* Center of Zero symbol Battery charge symbol Negative symbol Pound, Ounce, Pound:Ounce symbol Stable weight symbol Gram, kilogram symbol Preset Tare symbol* Pieces symbol* Pointer symbols* Percent symbol* Note: * Not Used...
EN-6 Valor™ 2000W Series 3.4 Initial Calibration When the scale is first installed it should be calibrated to ensure accurate results. Before performing the calibration, be sure to have the appropriate calibration weights as listed in table 3-3. Press and hold Cal for 5 seconds, the display will show [Cal]. When the button is released, the display will show [SpaN].
Page 8
Valor™ 2000W Series EN-7 4. OPERATION – V22XW 4.1 Controls Figure 4-1. Front Control Panel Valor 2000W – V22XW TABLE 4-1 Button Functions Short Press (when on): Sets display to zero Short Press (when off): Turns scale on Long Press (when on): Turns the scale off Short Press (in Menu): Selects / accepts displayed setting Long Press: Toggles through active Units...
Page 9
EN-8 Valor™ 2000W Series LED Indicators Figure 4-2. Valor 2000W – V22XW Rear display. The colored LED indicators on the right side of the control panels are used in Checkweigh mode (section 4.5) and will light up according to the following rules: (Red) Loads >...
Valor™ 2000W Series EN-9 4.4 Weigh Mode Press and hold Mode until [wJeIGH] (Weigh) is displayed. If required, place an empty container on the pan and press Tare. Add sample to the pan or container. The display shows the weight of the sample.
EN-10 Valor™ 2000W Series 4.6 Menu Settings The User Menu allows the customizing of scale settings. Note: Valor 2000W – V22PW does not have these Menus. 4.6.1 Menu Navigation User Menu: Menu: C.A.L M.O.d.E U.n.i.t E.n.d Span Reset Check Menu Items: lb:oz To Enter the Menu Mode...
Page 12
Valor™ 2000W Series EN-11 4.6.4 Unit Menu Enter this menu to activate units so they will be accessible with the Units button. The units in the menu must be turned “on” to be active. Note: Available units vary by model and local regulations. 4.6.5 End Menu Press ‘Yes’...
EN-12 Valor™ 2000W Series 5. MAINTENANCE 5.1 Cleaning The housing may be cleaned with a cloth dampened with a mild detergent if necessary. Do not use solvents, chemicals, alcohol, ammonia or abrasives to clean the housing or control panels. 5.2 Cleaning the Plastic Pan 5.2.1 Uninstalling and cleaning the plastic pan Please follow the steps below to clean the plastic pan: 1.
Valor™ 2000W Series EN-13 5.3 Troubleshooting The following table lists common problems and possible causes and remedies. If the problem persists, contact OHAUS or your authorized dealer. TABLE 5-1 Symptom Possible Cause Remedy No power to scale Verify connections and power...
EN-14 Valor™ 2000W Series 6.1 Specifications TABLE 6-1 V22PWE1501T V22PWE3T V22PWE6T V22PWE15T V22PWE30T MODEL** V22XWE1501T V22XWE3T V22XWE6T V22XWE15T V22XWE30T 1.5 kg x 0.0002 kg 3 kg x 0.0005 kg 6 kg x 0.001 kg 15 kg x 0.002 kg 30 kg x 0.005 kg 1500 g x 0.2 g 3000 g x 0.5 g 6000 g x 1 g...
Page 16
Mark Standard This product conforms to the EMC Directive 2004/108/EC, the Low Voltage Directive 2006/95/EC. The Declaration of Conformity is available online at europe.ohaus.com/europe/en/home/support/compliance.aspx. AS/NZS CISPR 11 CAN/CSA-C22.2 No. 61010-1-12 UL Std. No. 61010-1 (3 edition) NSF/ANSI 169–2009...
During the warranty period OHAUS will repair, or, at its option, replace any component(s) that proves to be defective at no charge, provided that the product is returned, freight prepaid, to OHAUS.