Keysight 85134E/F/H NMD-2.4 mm -f- to 3.5 mm and 85135E/F NMD-2.4 mm -f- to 7 mm Flexible Test Port Return Cables Operating and Service Manual...
Page 4
AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. KEYSIGHT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN.
Page 5
• Never coil the cable too tightly. • Never allow the cable to hang down from an instrument test port, or a storage fixture, or while holding it. • Never bend the cable an excessive number of times. 85134E/F/H & 85135E/F...
Page 6
Printing Copies of Documentation from the Web To print copies of documentation from the Web, download the PDF file from the Keysight web site: • Go to http://www.keysight.com. • Enter the document’s part number (located on the title page) in the Search box.
NMD-2.4 mm test ports. 85134E and 85135E The 85134E and 85135E sets contain a single cable. Use it when a DUT is connected, with the appropriate adapter, directly to the test set test port. See Figure 1-1.
Verify that the shipment is complete by referring to Table 4-1 on page 4-4. If the case or any device appears damaged, or if the shipment is incomplete, notify Keysight. “Contacting Keysight” on page 4-3. Keysight will arrange for repair or replacement of incomplete or damaged shipments without waiting for a settlement from the transportation company.
(you can order a single cable rather than the entire cable set), and for recommended items not included with the cable set. Refer to “Contacting Keysight” on page 4-3 information about communicating with Keysight. 85134E/F/H & 85135E/F...
DC to 26.5 ≤0.31 √f + 0.2 85134F 85134H ≤1.3 ≥17.7 ≤0.46 √f + 0.3 85135E DC to 18 ≤0.31 √f + 0.2 85135F a. f = frequency in GHz. b. Center conductor shoulder behind outer conductor mating plane. 85134E/F/H & 85135E/F...
This condition will indicate a positive value on the connector gage. Recession is the condition in which the center conductor is set back from the outer conductor mating plane. This condition will indicate a negative value on the connector gage. 85134E/F/H & 85135E/F...
Page 16
-0.0001 to -0.0005 PSC-3.5 mm -f- 7 mm -0.005 to -0.021 -0.0002 to -0.0008 0.05 to 0.25 0.002 to 0.010 a. Center conductor shoulder behind outer conductor mating plane. b. With collet removed. c. With collet in place. 85134E/F/H & 85135E/F...
3. Connect the other end of the cable to the test port. 4. Remove the short from the cable. Refer to Chapter 4 for part numbers and instructions for ordering ESD protection devices. Figure 3-1. ESD Protection Setup 85134E/F/H & 85135E/F...
If a connector shows deep scratches or dents, particles clinging to the mating plane surfaces, or uneven wear, clean and inspect it again. Cables with damaged connectors should be repaired or discarded. Determine the cause of damage before connecting a new, undamaged connector in the same configuration. 85134E/F/H & 85135E/F...
The precision slotless connector was developed to eliminate this problem. The PSC has a center conductor with a solid cylindrical shell, the outside diameter of which does not change when mated. Instead, the center conductor has an internal contact that flexes to accept the male pin. 85134E/F/H & 85135E/F...
Clean the connector threads. c. Let the alcohol evaporate, then blow the threads dry with a gentle stream of clean, low-pressure compressed air or nitrogen. Always completely dry a connector before you reassemble or use it. 85134E/F/H & 85135E/F...
Page 22
Let the alcohol evaporate, then blow the connector dry with a gentle stream of clean, low-pressure compressed air or nitrogen. Always completely dry a connector before you reassemble or use it. 4. Reinspect Inspect the connector again to make sure that no particles or residue are present. 85134E/F/H & 85135E/F...
Use, Maintenance, and Care of the Cables Gaging Connectors Gaging Connectors The gages available from Keysight Technologies are intended for preventive maintenance and troubleshooting purposes only. (See Table 4-1 on page 4-4 for part number information.) They are effective in detecting excessive center conductor protrusion or recession, and conductor damage on test cables and other accessories, DUTs, and calibration kit devices.
Refer to “Center Conductor Pin Depth” on page for definitions of protrusion and recession. Figure 3-2 Reading the Connector Gage 85134E/F/H & 85135E/F...
7 mm gage: If the needle has moved clockwise, the center conductor is protruding by and amount indicated by the black numbers. If the needle has moved counterclockwise, the center conductor is recessed by an amount indicated by the red numbers. 85134E/F/H & 85135E/F...
Page 26
Compare the average reading with the observed pin depth limits in Table 2-3 on page 2-4. 3-10 85134E/F/H & 85135E/F...
Page 27
Use, Maintenance, and Care of the Cables Gaging Connectors Figure 3-3 Gaging 2.4 mm and 3.5 mm Connectors 85134E/F/H & 85135E/F 3-11...
Page 28
Use, Maintenance, and Care of the Cables Gaging Connectors Figure 3-4 Gaging 7 mm Connectors 3-12 85134E/F/H & 85135E/F...
6. Carefully align the connectors. As you make the actual connection, be sure the connectors align perfectly. 7. Push the connectors straight together. Do not twist or screw the connectors together. 85134E/F/H & 85135E/F 3-13...
Page 30
3-5. Wrenches opposing each other (greater than 90 degrees apart) will cause a lifting action which can misalign and stress the connections of the devices involved. This is especially true when several devices are connected together. 3-14 85134E/F/H & 85135E/F...
Page 31
Any give at all in the handle is sufficient torque. Connecting Precision 3.5 mm and SMA Connectors An SMA connector will mate with a precision 3.5 mm connector. Because SMA connector tolerances are less stringent than those of precision 3.5 mm connectors, gage the SMA 85134E/F/H & 85135E/F 3-15...
When you use cables to make a precise calibration, you may have to fixture the cables to prevent excessive movement after the calibration. In some cases, unless you restrict cable movement, you may not be able to perform a verification after the calibration, especially if you use a precision calibration kit. 3-16 85134E/F/H & 85135E/F...
Page 33
Use, Maintenance, and Care of the Cables Handling and Storage NOTE After you perform a calibration, move the cables as little as possible. Every time you bend a cable, the phase changes slightly. 85134E/F/H & 85135E/F 3-17...
Page 34
Use, Maintenance, and Care of the Cables Handling and Storage 3-18 85134E/F/H & 85135E/F...
To order a listed part, note the description, the part number, and the quantity desired. Telephone or send your order to Keysight Technologies (see “Contacting Keysight” on page 4-3).
Assistance with test and measurement needs and information on finding a local Keysight office are available on the Web at: http://www.keysight.com/find/assist If you do not have access to the Internet, please contact your Keysight field engineer. NOTE In any correspondence or telephone conversation, refer to the Keysight product by its model number and full serial number.
Page 38
Replaceable Parts Returning a Cable or Cable Set to Keysight Table 4-1 Replaceable Parts for the 85134E/F/H Cable Sets Keysight Part Description Cables 85134E Single Cable: NMD-2.4 mm –f– to PSC-3.5 mm –f– 85134-60006 85134F Cables: NMD-2.4 mm –f– to NMD-3.5 mm –m–...
Page 39
Replaceable Parts Returning a Cable or Cable Set to Keysight Table 4-1 Replaceable Parts for the 85134E/F/H Cable Sets Keysight Part Description 3.5 mm female gage set 85052-60043 10x Magnifying Glass 1000-1114 3.5 mm 50 Ω fixed termination -m- 00902-60003 3.5 mm 50 Ω...
Page 40
Replaceable Parts Returning a Cable or Cable Set to Keysight Table 4-2 Replaceable Parts for the 85135E/F Cable Sets Keysight Part Description Cables 85135E Single Cable: NMD-2.4 mm –f– to 7mm 85135-60004 85135F Cables: NMD-2.4 mm –f– to 7 mm...
Page 41
Replaceable Parts Returning a Cable or Cable Set to Keysight Table 4-2 Replaceable Parts for the 85135E/F Cable Sets Keysight Part Description Collet extractor tool 5060-0370 7 mm Airline (10 cm) 85051-60007 “Clarifying the Terminology of a Connector Interface” on page 1-3 a.
Page 42
Replaceable Parts Returning a Cable or Cable Set to Keysight 85134E/F/H & 85135E/F...
Prior to making connections to the network analyzer, carefully review the connector care information provided with your product. See the following table for quick reference tips about connector care. 85134E/F/H & 85135E/F...
Page 45
Use a torque wrench for final connect Tighten past torque wrench “break” point a. Use isopropyl alcohol in a well-ventilated area, allowing adequate time for moist alcohol to evaporate and fumes to disperse prior to energizing equipment. 85134E/F/H & 85135E/F...
Page 46
Connector Care Quick Reference Principles of Microwave Connector Care 85134E/F/H & 85135E/F...
Page 47
NMD type connector precision 3.5 mm and SMA 3-16 dial PSC type handling quick reference master reading slotless zeroing SMA and precision 3.5 mm 3-16 gage master threads part numbers type 85134E/F/H & 85135E/F Index-1...
Page 48
NMD connectors frequency frequency range humidity insertion loss open-end wrench 3-16 return loss part number temperature ordering parts torque wrench 3-14 stability, phase static discharge parts storage 3-16 numbers altitude ordering humidity Index- 85134E/F/H & 85135E/F...
Page 49
3-14 visual inspection warranty, documentation wrench open-end 3-15 3-16 part numbers proper positioning of 3-15 torque 3-14 3-15 precautions for use of 3-15 proper use of 3-15 zeroing connector gage 85134E/F/H & 85135E/F Index-3...