  • Page 1 Keysight 11923A 1.0 mm Connector Launch Assembly Operating and Service Manual...
  • Page 2 Declarations of Conformity for this product ceed beyond a CAUTION notice until the indi- DFARS and are set forth specifically in writing and for other Keysight products may be down- cated conditions are fully understood and met. elsewhere in the EULA. Keysight shall be loaded from the Web.
  • Page 3: Certification

    Certification Keysight Technologies certifies that this product met its published specifications at the time of shipment from the factory. Keysight Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology (NIST, formerly NBS), to the extend allowed by the Institute’s calibration facility, and to the calibration facilities of the other International Standards Organization members.
  • Page 4: Waste Electrical And Electronic Equipment (Weee) Directive

    To return this unwanted instrument, contact your nearest Keysight Service Center, or visit http:// about.keysight.com/en/companyinfo/environment/takeback.shtml for more information. Sales and Technical Support To contact Keysight for sales and technical support, refer to the support links on the following Keysight websites: – www.keysight.com/find/adapters (product-specific information and support, software and documentation updates) –...
  • Page 5: Table Of Contents

    ............30 Keysight 11923A Operating and Service Manual...
  • Page 6 ..........37 Keysight 11923A Operating and Service Manual...
  • Page 7: List Of Figures

    ........33 Keysight 11923A Operating and Service Manual...
  • Page 8 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. Keysight 11923A Operating and Service Manual...
  • Page 9: List Of Tables

    ....... . . 35 Keysight 11923A Operating and Service Manual...
  • Page 10 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. Keysight 11923A Operating and Service Manual...
  • Page 11 Keysight 11923A 1.0 mm Connector Launch Assembly Operating and Service Manual General Information Connector Launch Overview...
  • Page 12: General Information

    Figure 1-1 Connector launch assembly The 11923A 1.0-mm female connector launch assembly is designed for ultra-high frequency (up to 110 GHz) coaxial signal transmission into a microstrip package. The launch assembly facilitates designing, building, testing, and troubleshooting high-frequency microcircuits, components, and custom prototypes.
  • Page 13: Incoming Inspection

    General Information Incoming inspection Keysight Technologies will arrange for repair or replacement of incomplete or damaged shipments without waiting for a settlement from the transportation company. When you send a device to Keysight Technologies, include the following information: – Your company name and address.
  • Page 14 General Information THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. Keysight 11923A Operating and Service Manual...
  • Page 15 Keysight 11923A 1.0 mm Connector Launch Assembly Operating and Service Manual Specifications Electrical Specifications Supplemental Characteristics Environmental Requirements...
  • Page 16: Specifications

    Specifications Electrical Specifications Electrical specifications are performance standards or limits against which the launch assemblies are tested. Table 2-1 Electrical specifications Frequency Value Insertion loss DC to 110 GHz -10 dB Keysight 11923A Operating and Service Manual...
  • Page 17: Supplemental Characteristics

    Supplemental Characteristics Supplemental characteristics are values which are typically met by the majority of the launches tested at Keysight Technologies. These supplemental characteristics are intended to provide useful information. They are typical but non-warranted performance parameters. Center conductor protrusion and pin depth The supplemental mechanical characteristics such as center conductor protrusion and pin depth are important supplemental characteristics related to electrical performance.
  • Page 18 0 (flush) to 0.025 mm (maximum recession) Launch pin extension 0.32 ± 0.05 mm Flatness of reference plane 0.010 mm (worst case) Flatness of end of launch body 0.010 mm (worst case) [a] Reference place to reference plane Keysight 11923A Operating and Service Manual...
  • Page 19: Environmental Requirements

    The operating temperature is a critical factor in the performance during measurements and between calibrations. Remember your fingers are a heat source, so avoid handling the devices unnecessarily. CAUTION Keysight 11923A Operating and Service Manual...
  • Page 20 Specifications THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. Keysight 11923A Operating and Service Manual...
  • Page 21 Keysight 11923A 1.0 mm Connector Launch Assembly Operating and Service Manual Mounting the Launch Mounting the Launch...
  • Page 22: Mounting The Launch

    – The centerline offset between the threaded and the package floor is calculated as follows: Offset = substrate thickness + 0.081 (radius of pin) + 0.075 (clearance for solder flow) – Gold plating is recommended for conductivity and corrosion protection. Keysight 11923A Operating and Service Manual...
  • Page 23: Threading The Launch Into The Package Or Fixture

    This is enough to create a slight deformation where the launch contacts the package floor to assure a good ground connection. Figure 3-2 Threading the launch into the package or fixture [1] This torque wrench is not available from Keysight Technologies. Keysight 11923A Operating and Service Manual...
  • Page 24: Mounting The Circuit Inside The Package

    If solder is used, it is important to control the volume of the solder flow to match the size of the NOTE launch pin and the signal trace. Be careful to avoid too much solder (capacitive) or too little (inductive). Figure 3-3 Circuit mounting detail Keysight 11923A Operating and Service Manual...
  • Page 25 Keysight 11923A 1.0 mm Connector Launch Assembly Operating and Service Manual Making Connections Making Connections Visual Inspection Pin Depth Connector Alignment Cleaning Connectors Connection Procedure Using the Torque Wrench Disconnection Procedure Handling and Storage...
  • Page 26: Making Connections

    Connector cleaning supplies and electrostatic discharge safety supplies are not provided but can be ordered separately. For cleaning and inspection, a 10X magnifying glass is available. Keysight 11923A Operating and Service Manual...
  • Page 27: Visual Inspection

    Figure 4-1 for visual guidelines to aid you in evaluating the contact integrity of a connector. Take careful note of the location of the cross hairs in relationship to the center of the following figures. Keysight 11923A Operating and Service Manual...
  • Page 28: Concentricity

    Making Connections Figure 4-1 Contact integrity Concentricity Figure 4-2 Figure 4-3 show the concentricity of both the male and female 1.0-mm connectors. Figure 4-2 Concentricity - female connector Figure 4-3 Concentricity - male connector Keysight 11923A Operating and Service Manual...
  • Page 29: Mating Plane Surfaces

    Connector wear eventually degrades performance. The more use a connector gets, the faster it wears and degrades. The wear is greatly accelerated when connectors are not kept clean. When your connectors become worn, replace them. Keysight 11923A Operating and Service Manual...
  • Page 30: Pin Depth

    The electrical performance of the connector depends, to some extent, on its pin depth. The electrical specifications take into account the effect of pin depth on the performance of the connector. Figure 4-4 shows a visual representation of proper pin depth (slightly recessed). Figure 4-4 Connector pin depth Keysight 11923A Operating and Service Manual...
  • Page 31: Connector Alignment

    (bending) force. 5 Use an open-end wrench to keep the device from rotating when you make a final connection with the torque wrench. Figure 4-5 Alignment Keysight 11923A Operating and Service Manual...
  • Page 32: Connector Misalignment

    Making Connections Connector misalignment Forced misalignment could damage the female center conductor. Figure 4-6 Forced misalignment To achieve misalignment as shown in Figure 4-6 you must use excessive force. NOTE Keysight 11923A Operating and Service Manual...
  • Page 33: Cleaning Connectors

    4 Gently wipe connecting surfaces with the end of the cleaning swab (refer to Figure 4-7). 5 Blow dry with a clean supply of compressed air. 6 Inspect and repeat cleaning procedure if necessary. Figure 4-7 Cleaning illustration Keysight 11923A Operating and Service Manual...
  • Page 34: Connection Procedure

    Use very light finger pressure (no more than 2 in-lbs of torque). 7 Relieve any side pressure on the connection from long or heavy devices or cables. This assures consistent torque (refer to “Using the Torque Wrench” on page 35). Keysight 11923A Operating and Service Manual...
  • Page 35: Using The Torque Wrench

    1.0-mm male nut while holding the hex portion of the launch with a 7-mm open-end backup wrench. – To make other 1.0-mm connections, use a 6-mm torque wrench and 6-mm backup wrench (refer to “Connection Procedure” on page 34). Keysight 11923A Operating and Service Manual...
  • Page 36: Disconnection Procedure

    Do not twist one connector out of the other, (like removing a light bulb). Turn the connector nut, CAUTION not the device body. Major damage to the center conductor and the outer conductor can occur if the device body is twisted. 4 Pull the connectors straight apart without twisting or bending. Keysight 11923A Operating and Service Manual...
  • Page 37: Handling And Storage

    – Do not set connectors contact-end down on a hard surface. The plating and the mating plane surfaces can be damaged if the interface comes in contact with any hard surface. Keysight 11923A Operating and Service Manual...
  • Page 38 Making Connections THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. Keysight 11923A Operating and Service Manual...
  • Page 39 This information is subject to change without notice. Always refer to the Keysight website for the latest revision. © Keysight Technologies 1999 - 2019 Edition 2, February 27, 2019 Printed in Malaysia *11923-90001* 11923-90001 www.keysight.com...

Table of Contents