  • Page 1 Keysight 16048A Test Leads Operation and Service Manual...
  • Page 2 EULA. governed by United States and Declaration of Conformity Keysight shall be under no obligation international copyright laws. to update, revise or otherwise modify Trademark Acknowledgments Declarations of Conformity for this the Software.
  • Page 3: Table Of Contents

    Maintenance ..............9 Keysight 16048A Test Leads...
  • Page 4 Contents Keysight 16048A Test Leads...
  • Page 5: Operation

    BNC (m) terminated-coaxial test leads. These test leads are used to attach user-fabricated test fixtures. DC bias levels of up to ±42V can be applied to the 16048A. Cable length is 1 meter. The 16048A is shown in Figure 1-1.
  • Page 6: Specifications

    Operation Specifications Specifications Table 1-1 Specifications of the 16048A Function: For use with 4 terminal-pair LCR Meters and Impedance Analyzers. Connector Type: BNC male Cable Length: Approximately 1m Maximum Voltage: ± 42V peak max. (AC+DC) Frequency Range: DC to 30 MHz...
  • Page 7: Typical Characteristics

     0.02 ----- -     f is the measurement frequency in megahertz. The incremental errors calculated from the equation in the table for measurements at frequencies above 1 MHz are additive. Figure 1-2 Keysight 16048A Test Leads...
  • Page 8: Compensation For Fixture Residual Impedance Error

    Compensation for Fixture Residual Impedance Error Compensation for Fixture Residual Impedance Error The 16048A has inherent stray capacitance, residual inductance, and residual resistance that affect the accuracy of measured values. To compensate for, or negate, these residuals to minimize measurement error, the instrument’s Open/Short compensation procedure should be performed.
  • Page 9: Service

    RoHS support part backward incompatible with non-RoHS 16048A. Special handling is needed while using the RoHS replacement part on non-RoHS 16048A. The original support part number is replaced by the respective “RoHS Compliant Replacement Part”.
  • Page 10 Service Maintenance Figure 2-1 Parts Identification Keysight 16048A Test Leads...
  • Page 11 1400-0719 CABLE TIE 1400-3284 CABLE TIE (1400-3284) GROMMET CABLE BUSHING 2360-0115 SCREW 16048-60610 TEST LEAD 3050-0010 WASHER 16048-60610 TEST LEAD 16021-50021 CABLE CRAMP 16048-60610 TEST LEAD 16021-50022 CABLE CRAMP 16048-60610 TEST LEAD BOOT-BNC SLEEVE SLEEVE INSULATOR Keysight 16048A Test Leads...
  • Page 12: Washer

    WASHER 2950-0001 2950-0001 16032-10022 PLATE PLATE Set Change: 16032-10622 16032-60001 BNC BRACKET 16032-60071 CONNECTOR (31 thru 36) ASSEMBLY a. Replace all the associated parts marked with as they are mutually dependent. *. Not separately replaceable.Order 16048-60010. Keysight 16048A Test Leads...