Page 2
2 SLIT LAMP HS-5000 2X - ---- ----------· - ------ IMPORTANT NOTICE .&.wARNING Potential elcctromagnetic or other interference between medical equipmcnts and other devices being operated together in t he same environmental may expert an adverse iníluence on functioning the medical equipment.
Page 3
---- ----------- - - - --------SLIT LAMP HS - 5000 2X HUVITZ reserves the right to make changes in its products or product specifications at any time and without prior notice, and is not required to update lhis documentation to reflect such changes.
6 SLIT LAMP HS-5000 2X - ---- ----------· - ------ lntroduction 1.1. Outline ofthe instrument The device works like ali slit lamps: stereomicroscopic examination of the eye under the slit light. microscopic examination of the o cular fundus and posterior vitreous body (using hubry's lens).
---- ----------- - - - --------SLIT LAMP HS - 5000 2X Safety lnformation 2.1. lntroduction Safety is everyone' s responsibillty. The safe use ofthis equipment is largely dependen! upon the installer, usar, operator, and maintainar. 11 is imperative that personnal study and become familiar with this anlira manual befare attempling to install, use, claan,...
Protective eanh connected to conductivc pans of Class I cquipment far safety putposes. Hot surface. 8 SLIT LAMP HS-5000 2X - ---- ----------· - ------ 2.2. Safety Symbols The lnternational Electrotechnical Comm ission (IEC) has established a set of symbols...
Page 9
---- ----------- - - - --------SLIT LAMP HS - 5000 2X 9 MEDI AL EUlPME T WITH RESPECTTO ELECTRIC SHOCK L6060I-I FIRE 1ECHA ICAL HAZARDS /CSAC222 O LY ACCORDANCE WITJ-1 UL 60601- 0.601.1 1, AND CA /CSA C22.2 0.601.1 Disposal ofyour old applíance When thís crossed--out wheeled bin symbol is anached to a product il meaos the product ís covered by lhe European Dírective 2002/96/EC.
10 SLIT LAMP HS-5000 2X- - - - - - ------- -------- 2.3. Environment factors Avoid the following environments for operation or storage: Where the equipment is exposed to water vapor. Don 't opernte an equipment with a wet hand.
Page 11
----- ------- - - - - - ---�SLIT LAMP HS-5000 2X 11 6® · � material Where equipment exposed chemical explosive gas. � � ..Be careful not to be inserted dust, especially, metal. Don't disa. s semble product open.
12 SLIT LAMP HS-5000 2X- - - - - - ------- -------- 2.4. Safety Precautions This equipment has t>een developed and tested according to safety standards as well as national and international standards. This guarantees a very high degree of safety for this device.
Page 13
----- ------- - - - - - ---�SLIT LAMP HS-5000 2X and maintained according to this manual and Service Manual. The manufacturar is not liable far damage causad by unauthorized tampering with the device(s). Such tampering will forfeit any rights to claim under warranty.
14 SLIT LAMP HS-5000 2X- - - - - - ------- -------- Features & Operating Condition 3.1. Brief Description The device works like all slit lamps: stereomicroscopic examination of the eye under the sllt light. microscopic examination of tha ocular fundus and posterior vitreous body (using hubry's lens).
Page 15
----- ------- - - - - - ---�SLIT LAMP HS-5000 2X _8cAUTION Because prolongecl imense light exposure can dama ge the retina, the LL s e of the device for ocular exarnination should not be unnecessarily prolonged, and the brightness setting should not exceed whai is needed to provide clear visualization of the target structures.
16 SLIT LAMP HS-5000 2X- - - - - - ------- -------- Standard Components The instrument is delivered in package. Taking it out of the package, make sure that none of the following parts i s missi ng. • A table top(N.B. the table top isn't su pplied with the slit lamp far twin table) on...
18 SLIT LAMP HS-5000 2X- - - - - - ------- -------- Composition Fixation point Control lever for changing enlargement Removable eye-pieces Knob for clamping the brea th screen Breath screen Control handle for the slit rotation 90-0-90 Co nnecting plug of the proj e ctor lamp...
Page 19
----- ------- - - - - - ---�SLIT LAMP HS-5000 2X 26) Locking pins for chin rest paper 27) Chin rest 28) Knobs to change the slit width 29) Pointer for the eye positioning 30) Head rest 31) Microscope 32) Bulb feeding contacts...
----- ------- - - - - - ---�SLIT LAMP HS-5000 2X lnstallation NOTE The applim1ces are always supplied with excellent packaging lo be able to stand up to ali kinds of transport and standard storage conditions. Picase contact the suppliers if any irregularities are found in the appliance dueto transport a-fter un p acking 7.1.
Page 24
24 SLIT LAMP HS-5000 2X- - - - - - ------- -------- • Place the head res! module(59} and make sure that its holes frt together with !hose of !he table top. Tighten !he screw with the screw driver [ Figure 5. Assembllng the instrument I] •...
Page 25
----- ------- - - - - - ---�SLIT LAMP HS-5000 2X • Assemble An optical group of !he slit projector(49) on !he Base wi t h orthogonal movements( 10). Then lighten the Guide screw wi t h the screw driver(-) •...
Page 26
26 SLIT LAMP HS-5000 2X----- --------------- 1 Floure 6. Assemblino the lnstrument II J 26/46...
----- ------- - - - - - ---�SLIT LAMP HS-5000 2X Trial Operations • Assembla !he chin-rest paper(47) on the chin-rest(27). • Le! !he palien! sil down comfortable with his chin on the chin-rest and his forehead against the forehead-rest.
28 SLIT LAMP HS-5000 2X- - - - - - ------- -------- Main Features Principal operativa characteri s tics of the instrument: • Observation without settling or tiring and rapid change of enlargements. • Base with cross-wise movernent (whi c h can be stopped) commanded by a s ingle j oystick for all adjustments(x y z) •...
----- ------- - - - - - ---�SLIT LAMP HS-5000 2X 1 O. Ordinary Maintenance The repairs illustrated below be carried out with the power cable disconnected from lhe main supply plug. Picase contact the suppliers for any other problems that cannot be resolved by the operations described below.
Page 30
30 SLIT LAMP HS-5000 2X- - - - - - ------- -------- touch !he bulb whlt fingers) • Fasten the bulb with the knob • Make sure that the stems of the bulb are completely inserted; insert the plug in !he main supply socket [ Figure 7.
----- ------- - - - - - ---�SLIT LAMP HS-5000 2X 10.2. Protection fuses replacement To replace the fuses proceed as follows: • The protection fuses are placed on the back of the SMPS, and insertad power socket. Befare doing anything else, disconnect the plug from the main supply socket •...
32 SLIT LAMP HS-5000 2X- - - - - - ------- -------- 10.3. Cleaning & Disinfection • To keep the device clean, apply clinical alcohol and cidex or cid ex plus on a soft cloth and scrub the body and the probe once a month. Do not use lacquer, thinner, ethylene, or the oxidizing substance.
----- ------- - - - - - ---�SLIT LAMP HS-5000 2X 11. EMC lnformation 11.1. Guidance and manufacturer's declaration - electromagnetic emissions The EUT is intended for use in the el e ctromagnetic environment specified The customer or !he user of the EUT shoul d assure that lt is used in such an environmenl.
34 SLIT LAMP HS-5000 2X- - - - - - ------- -------- 11.2. Guidance and manufacturer's declaration - electromagnetic immunity The EUT is intended far use in the eleclromagnetic environment specified below. The customer or the user of the EUT should assure that it is used in such an environment.
Page 35
----- ------- - - - - - ---�SLIT LAMP HS-5000 2X <5% UT EUT image inlensifi e r be ( < 9 5 % for 5 s powered from uninterruplible power for 5 s supply ora battery. Power frequency 3Alm...
36 SLIT LAMP HS-5000 2X--- - - - ------- -------- 11.3. Guidance and manufacturer's declaration - electromagnetic immunity The EUT is intended for use in the electromagnelic environment specified below. The customer or the user of the EUT should assure that it is used in such an environment.
Page 37
----- ------- - - - - - ---�SLIT LAMP HS-5000 2X and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as dete � -mined by an electromagnetic site survey, should be less than the compliance level in each frequency range.
38 SLIT LAMP HS-5000 2X- - - - - - ------- -------- 11.4. Recommended separation distances between portable and mobile RF communications equipment and the EUT The is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The custo mer or the user of the EUT can help Preven! electromagnetic inteñerence by maintaining a mínimum distance between portable and mobile RF communications equipment (transmitters) and the EUT as recommended b elow, according to the maximum output power of the communications...
40 SLIT LAMP HS-5000 2X- - - - - - ------- -------- 11.6. Guidance and manufacturer's declaration - electromagnetic immunity The EUT is intended for use in the electromagnelic environment specified below. The cust omer or the user of the EUT should assure that it is used in such an electromagnetic environment.
Page 41
----- ------- - - - - - ---�SLIT LAMP HS-5000 2X 41 NOTE 1) These guidelines may not apply in all situations. Electromagnetic propagation is affected by ab sorption and reflection from structures, objects and people. NOTE 2) lt is essential that the actual shielding effectiveness and filler attenuation of the shielded location be verified to assure that they meet the mínimum specification.
----- ------- - - - - - ---�SLIT LAMP HS-5000 2X 43 13. Technical Specifications MICAOSCOPE HS-50002X Type Galileian converging binocular Magnification 2 positions rotatlon drum Eyepiece 12.5x Total magnification 10x, l 6x Real field view 38.5, 22.2, 15.2, 10.5, 6.7 mm...
Page 44
44 SLIT LAMP HS-5000 2X- - - - - - ------- -------- BASE Vertical Movement Min. 28mm Lon g itudinal Movement Min. 78mm. Lateral Movement Min. 98mm Fine base Movement Min. 10mm 100-240 V AC 50 / 60HZ 2.0A Power su pp l y...
----- ------- - - - - - ---�SLIT LAMP HS-5000 2X 45 14. Service lnformation How to contact service: Ifthere are any problems with the equipment. picase follow the steps below: • First ali, re fer to the I O. Maintenance and Troubleshooting sections according to the problem th at you are encountered.