7.
Adjust 1 microsecond period capacitor
C115F
a.
ot the Type 114 PERIOD control to 1 µ , the WIDTH
control to 100 n , and both PERIOD
ABLE controls to CAL (see Fig. 4-8).
b.
et the Type 547 TIME/CM (B ) switch to 1 µ EC. Push
in the PULL TO INVERT switch on Channel 2 of the Type
1A1.
c.
et the Type 180A to deliver 1 microsecond markers .
d. Using the Type 1A1 PO ITION controls, put the peak
of the first pulse from the Type 114 to some reference point
on the first pulse from the Type 180A (see Fig. 4-9).
e. Adjust capacitor C115F (see Fig. 4-10) so that the peaks
of the pulses from the Type 114 occur at the same point on
each succeeding pulse from the Type 180A as shown in Fig.
4-9.
Allowable error is 3%, but it is possible to adjust
much closer than this.
Fig
4-8. Control settings for adjusting the 1 microsecond period capacitor .
and WIDTH — VARI ­
Maintenance and Calibration — Type 114
8.
Check period accuracy, 10 mS to 1
a.
et th e Type 114 PERIOD control to 10 m
WIDTH control to 10 µ .
Fig. 4-10
Location of
1
microsecond period capacitor C115F.
b. Turn the Type 1A1 MODE switch to ALT and pull out
the Channel 2 PULL TO INVERT switch
c.
et the Type 180A to deliver 10 millisecond markers.
d.
et the Type 547 TIME/CM |B) switch to lOm EC and
adjust the TRIGGERING LEVEL control for a stable display.
e. Check for horizontal alignment between the two sets
of markers (see Fig. 4-7).
Horizontal alignment should be
within 3%.