Troubleshooting

The detector fails to turn on:
Batteries may be too weak to enable normal operation. Change batteries.
Under harsh environmental conditions, the pole connector in the battery compartment
door may have become oxidized. Clean off any oxide and try again.
The detector won't respond to a known leak
The sensor tip may be contaminated.
Clean as pointed out above.
The sensor tip is subject to aging and may be worn. Use a new sensor tip.
The detector gives false alarms
The ambient atmosphere's content of detectable substances may have changed Press
RESET to suppress any background levels that may be present.
The sensor may be contaminated or worn out.
Clean sensor as described above, or use new sensor.
The sensor was brought in contact with a metallic surface.
Press RESET while in contact with the specimen to readjust the detector.
The air exchange in the sensor was obstructed.
Always approach surfaces at an angle, never flat on. Press RESET to readjust the detector.