The SF6 LEAKMATE is a highly sensitive, processor controlled SF
based on the proven detection principle of Negative Corona Discharge. It is small, lightweight, and
easy to operate even for untrained personnel. With its audible and visual alarm indications and
adjustable sensitivity, it is optimally suited for not only finding but also pinpointing SF
to approximately 1E-6 cc/s.
Warnings
In confined spaces, a severe SF
deprived to a dangerously low level. Entering such confined spaces poses serious danger of
asphyxiation. Always ventilate such rooms before entering.
Do not operate the instrument with no sensor tip in place; there is risk of electric shock.
Commissioning

3.1. Battery Installation

Hold the instrument with the probe facing downwards, and the front panel towards you. With a
firm press by thumb, slide the battery compartment door towards the back of the instrument, and
remove it.
Batteries are inserted with the positive cap facing outwards, as indicated by the marking visible
inside the battery compartment.
3.2. Installing / Replacing the Sensor Tip
WARNING:
Do not power up the instrument when no sensor tip is in place. Doing so may pose
hazard of electric shock.
Make sure the instrument is powered down before removing the sensor tip.
The sensor tip is installed by gently screwing it on hand-tight.
Only use sensor tips that are specified for this instrument. Using any other type of sensor tips,
even if they may mechanically fit, will result in compromised detection performance.
Unrivalled Gas
leak can flood the room to a concentration where oxygen is
6
Detection.
SF6 Leakmate Instrument User Manual V1.0
leak detector for mobile use,
6
Page 4 of 9
leaks down