Note:
Petroleum jelly may be applied prior to lancing the
site. This can help the blood to form into a droplet and avoid
spilling into the hair.
Step 9
After the
symbol appears on the screen, apply the blood
sample to the narrow end of the test strip till the meter
beeps. If the confirmation window is not filled in time because
of abnormal viscosity (thickness and stickiness) or insufficient
volume, the Er4 message may appear. It is recommended to
place the test strip vertically into the blood sample site as
shown below.
28 www.i-sens.com
Good Sample
Insufficient
Sample
Note:
The meter may switch off if the blood sample is not
applied within 2 minutes of the
screen. If the meter turns off, remove the strip and reinsert it,
and apply the blood sample after the
the screen.
Step 10
The test result will appear after the
meter counts down from 5 to 1. The
result will be automatically stored in
the meter's memory.
Gently apply pressure to the test site
with sterile gauze or cotton wool to help
stop the blood flow.
Step 11
Slide the ejection button to discard
the used strips safely in disposable
container.
If the test strip is removed after the
test result is displayed, the meter
will automatically switch off after 3
seconds.