screen displays the message «Lb» (Low battery), the
handpiece beeps 4 times, and the start of a curing
cycle will be inhibited.
• You must then replace the ScanWave on its charger.
1.6 r
echarging
The translucent «Neck» of the charger (fig. 1-10) shows
the charger operation status and the level of lighting
power emitted by the light.
Place the charger on a hard, stable surface with a
flatness of less than 5°, then plug the connector into
the grid supply. The translucent part of the support
successively emits three flashes: red, green and blue.
Then insert the light guide in the charger and ensure
charging terminals are in contact with battery
connections.
The translucent part then emits a blue light signalling
that the battery is charging. The blue light goes out
when charging is complete.
1.7 c
hecking lighting Power
The charger can be used to check the lighting power
emitted by ScanWave. To do this, a control window is
located in the rear section of the support (fig. 1-11).
To check the lighting power emitted:
• Select Scan mode; continuous; 9 seconds.
• Clean the optical guide, make sure that it is not
damaged, and then insert it into the handpiece.
• Also make sure that the control window is unbroken
and clean.
• Place the tip of the optical guide flat on the control
window and activate ScanWave.
• If the transluscent «Neck» of the charger lights up
green, the lighting power is sufficient. If the lighting
power is not sufficient, this will be signalled by a
fixed red or flashing red/green light. In this case,
refer to chapter X.
1.8 t
Model name
Manufacturer
Medical category
Handpiece
Weight
Dimensions
Operating cycle
Category
Protection
Protection index
AC power adapter
Input voltage
Frequency
Output voltage
Output current
Category
Protection index
Charging base
Input voltage
Protection
Operating cycle
Protection index
Battery
Type
Dimensions
Capacity
Optical specifications
LED sources
Wavelength range
Central wavelength
Intensity
1500 mW/cm² ± 10% (for a 7.5mm fiber : active
diameter 6.8)
2200 mW/cm² ± 10% (for a 5.5mm fiber : active
diameter 4.8)
Maximum exposure time
60 seconds (pulse mode at
Laser aiming system
Typical wavelength
User's Manual ScanWave ● J05451 ● V5 ● (13) ● 05/2013 ● NG39EN020E ● 5
echnical sPecifications
ScanWave
SATELEC
IIa
directive 93/42/CEE
185g
Ø34 x 220mm
See II - Routine operation -
Clinical care
Type B
1,5 A T Fu1 fuse (non
accessible) 125V
IPX0
100V AC to 240V AC
50Hz to 60Hz
12 V DC
0.8 A
II
IP 41
12V DC
3 A T Fu1 fuse (non
accessible) 125V
Continuous operation
IPX0
Lithium-Ion
Ø34 x 115mm
2500mAh
4
390nm-505nm
405nm / 440nm / 460nm
and 480nm
655nm
according
to
the
1500 mW/cm² ± 10%)