an icon indicates the type of light emission selected:
continuous, flashing or gradual.
• Place the optical guide as close as possible to the
surface of the equipment to be photocured but
avoid contact as this may reduce the quality of the
curing process. The use of a transparent strip is
recommended.
• Press the ON/OFF button to start the polymerization
cycle.
• A countdown counter on the LCD screen shows the
time remaining.
• Press the button again for longer to stop the current
cycle at any time.
III – CONFIGURATION
7.1 l
aser sight
ScanWave has a laser guiding system which positions
the light beam in the direction of the material to be
photocured.
At the start of a cycle, press and hold down the ON/
OFF button or the trigger to display a red sight circle.
Release them to make the sight circle disappear and
activate the light. This function allows you to view the
light zone associated with optimum power. The laser
radiation passes into the optical guide. Do not look in
the optical guide.
3.2 c
ycle initiation
By default, in order to ensure that the device can
be used easily by everyone, cycles are initiated by
pressing the ON/OFF button or by pushing the control
trigger.
However, depending on the required ergonomics, the
cycle start control can be configured:
• One long press on the Cycle C button makes the
following screen appear:
• Activation is then possible either by pressing the
button or the control trigger. Bo signifies «Both».
• A short press on the menu C button allows you to
select the pen-style hold. Initiation is then only
possible by pressing the ON/OFF button and the
screen displays PE: «Pen-Style».
• To validate the modification and exit configuration
mode, press the Mode M button again.
• However, if you prefer the gun-type hold, quickly
press the Cycle C button. Launch is then only possible
by pressing the control trigger and the screen displays
GU: «Gun-type».
• To validate the modification and exit configuration
mode, press the Mode M button.
• Proceed in the same way to change the initiation
mode again.
3.3 s
After 3 minutes without use, the unit goes into standby
mode (low consumption), the LCD screen backlighting
switches off.
The unit can be taken out of standby mode by simply
pressing one of the buttons.
IV – SECURITIES
ScanWave is equipped with a system for the detection
of possible unit operation anomalies.
4.1 e
During intensive use, an excessive temperature may be
detected and the screen then displays the letters «OH»
(Over Heat) and the audible warning (beep) sounds 4
times. The light can now not be switched on.