Error Message
Supply Pressure Too Low
Value out of range
Function not available
Illegal Operation
Find Quality Products Online at:
Table 7. Error Messages (cont.)
Explanation
Failed to create supply pressure for the
current setpoint.
Value input is out of range.
Selected function is not available for the
current state.
Illegal operation for the current state.
GlobalTestSupply
www.
Automatic Pressure Calibrator
Potential Root Cause
1. The pressure volume is too large, reduce
the volume and try again.
2. The leakage of pressure volume is too
large. Test the leakage and try again.
3. Ambient pressure is lower than the negative
gauge setpoint. Change the negative gauge
setpoint to a reasonable value at current
ambient pressure.
4. The unit needs service.
Try to configure a valid value.