Error Message
Lost connection to HART
transmitter, Task aborted
Lost connection to FC devices,
Task aborted
Lost connection to RTD, Task
aborted
Open circuit detected
Invalid Task Setting
Find Quality Products Online at:
Table 7. Error Messages (cont.)
Explanation
Running task aborted, due to lost
connection to HART transmitter.
Running task aborted, due to lost
connection to Fluke Connect
secondary unit.
Running task aborted because RTD
was removed.
Open circuit detected while sourcing
mA.
Task setting contains invalid items.
GlobalTestSupply
www.
Automatic Pressure Calibrator
Potential Root Cause
Make sure the transmitter is connected and
powered. Try HART polling again. After
connected, rerun the task.
Turn on the FC secondary device and make
sure FC is active. Redo the discovery
process. After connected, run the task again.
Make sure the RTD probe is connected and
run the task again.
Connect the circuit properly and try again.
Verify zero/full-scale value defined in the task
are in the range of the selected pressure
source and try again.