12.2 Optionale external charger CHJ2

The charger can be mounted either between the control box and the battery on the
patient hoist or separately on a wall or elsewhere. If the charger is mounted between
the control box and the battery, the functions of the control box are deactivated when
mains power is connected to the charger.
The charger has the following properties:
• It is mains charger with an input voltage of 100-240 V AC
• Charging time: max. 4 hours for the BAJ1/BAJ2 battery set
• Replaceable power cable with retaining ring
• Green light ON illuminating: Charger is connected to mains power
• Yellow light CHARGING illuminating: Battery is charging Lights OFF means: the
• battery is fully charged
• Simple assembly of charging device on JUMBO mount (optional)
• Unless the EMERGENCY STOP button has been activated the charging process takes
place
Charger is connected to
mains power, battery
is charging
Plug locks in place
Plug socket