Warning Lights; Cbjh; Chjc Maintenance Warning Light Flashes - NOVACARE ProLift Series Instructions For Use Manual

Patient hoist soloution made of steel
Table of Contents

Advertisement

17

11. Warning lights

11.1 CBJH

Battery: if the battery capacity is low, a continuous alarm signal will sound when the
manual controller is operated.
Overload: if an excessive load is exerted and the CBJH drive unit is overloaded, two
audible signals will sound (power limit exceeded). The power is disconnected and the
drive unit will stop. The drive unit can be restarted as soon as the overload has been
remedied.

11.2 CHJC Maintenance warning light flashes:

Every 12 months (in line with the recommendation contained in
EN10535) / after 8,000 cycles, whichever occurs first.
The first time the maintenance symbol lights up, the control box
produces an audible warning. Once the maintenance date is
reached, the maintenance warning light remains on for 2 minutes
after use.
The control box then switches off, in order to conserve the battery.
When the maintenance warning light is illuminated, the system
can still be operated and will work normally. Contact us or an
authorised service engineer. Compliance with the maintenance
intervals serves to protect the patients and users.
Battery:
3 LEDs:
- Green: the battery need not be charged
- Yellow: the battery requires charging
- Yellow: the battery requires charging,
audible signal also produced when manual controller activated
If hoist continues to be used on low battery charge, the LED will flash and will produce
a continuous audible signal. At this point, it will still be possible to lower the hoist arm
(lifting and trolley base adjustment will no longer be possible).