12. Batteries (charging/replacing)

Attention!
The batteries must be replaced after four years at the latest - possibly sooner,
depending on extent of use. Frequent and rapid discharge of the batteries reduces
their service life. For an optimal service life, the batteries should be connected to
mains power as frequently as possible. The batteries must be charged at least once
every three months, or they will be damaged by the selfdischarge process. The CBJC
electronic controller contains an electronic control box and a separate battery box.
Fully charge before first use!
Akku vor erstem
Gebrauch vollständig
laden!

12.1 CBJ-Home:

Fully charge before
first use!
¡Cargue la batería
antes de usarla por
primera vez!
Charger complète-
1
Charging light (yellow)
ment la batterie
1
avant premier usage !
2
LED showing manual controller activated (green)
2
Button for emergency electrical lowering
3
3
1. Charging can only take place if the emergency „Stop" button
2. The indicator (1) flashes during charging.
3. The charging process is completed once both LEDs (yellow and
Battery box
Battery box
has been deactivated.
green) remain illuminated.