3.7.6 0x2207 – The Pressure Calibration Gain
Refer to "0x2200 – To Change the Calibration Data" on page 8. Use 0x2207 (Pressure Gain) to
apply a correction to the compensated source pressure (refer to Chapter 4., "Maintenance,"
on page 15).
Compensated
source
pressure
3.7.7 0x2208 – The Pressure Calibration Offset
Refer to "0x2200 – To Change the Calibration Data" on page 8. Use 0x2208 (Pressure Offset)
to apply a correction to the compensated source pressure (refer to Chapter 4.,
"Maintenance," on page 15).
Compensated
source
pressure
3.7.8 0x2209 – The Temperature Calibration Gain
Refer to "0x2200 – To Change the Calibration Data" on page 8. Use 0x2209 (Temperature
Gain) to apply a correction to the compensated source temperature.
Compensated
source
temperature
© 2018 Baker Hughes, a GE company – All rights reserved.
To Change the Operation – Objects: 0x2000 – 0x2FFFF
Object 0x2207:
Pressure Gain
x Gain
correction
Figure 3-2: Pressure Calibration Gain
Object 0x2208:
Figure 3-3: Pressure Calibration Offset
Object 0x2209:
Temperature Gain
x Gain
correction (°C)
Figure 3-4: Temperature Calibration Gain
Object 0x6100 01
Pressure Value
Field Value (mbar)
Object 0x6100 01
Pressure Value
Object 0x6130 02:
Temperature Value
Process Value (°C)