3.6.3 0x1010 01 – To Save Changes to the Data Dictionary
Use object 0x1010 01 (Store Parameter Field) to save the data dictionary to non-volatile
storage.
Example:
Set the value to 0x65766173 = evas (in ASCII)
Note: This does not overwrite the factory data and the sensor can be returned to its
factory state by using 0x1011 as described below.
3.6.4 0x1011 01 – To Re-apply the Factory Values
Use object 0x1011 01 (Restore Default Parameters) to re-apply all the factory values.
Example:
Set the value to 0x64616F6C = daol (in ASCII)
Having re-applied the factory values, the sensor should be power cycled for the values to be
applied.
3.6.5 0x1800 02 – The PDO Transmission (Type or Period)
Use object 0x1800 02 (Transmission Type) to change the type of PDO transmission or the
period between each PDO transmission (refer to Appendix C., "Bibliography," on page 55).
3.6.6 0x1A00 – The Data in the 'Transmit PDO'
Use object 0x1A00 (Transmit PDO Mapping) to change the data sent in the 'Transmit PDO'
(maximum size = 8 bytes).
To change the transmit PDO, set bit 31 of 0x 1800 1 "COB-ID" to 1 and set 0x IA00 1 to 0.
After changing the PDO the process should be reversed.
Example:
To monitor the current status of the pressure value, set Sub-index 0x02 to
0x61500108 = Object 0x6150 01, 1 byte of data.
Status
0
1
2

3.6.7 0x1F80 - NMT Start-up

Use object 0x1F80 to set the start-up mode of the sensor.
Status
0
© 2018 Baker Hughes, a GE company – All rights reserved.
To Change the Operation - Objects: 0x1000 - 0x1FFFF
Description
The value is in the limits of 0x6148 and 0x6149.
The value is more than the limit of 0x6149.
The value is less than the limit of 0x6148.
Description
Sensor power-up in pre-operational mode.
DPS 5000 Series Instruction Manual-English | 7