7. INFORMATION RELATING TO ELECTROMAGNETIC
COMPATIBILITY IN ACCORDANCE WITH IEC 600601-1-2
The device is designed and manufactured in compliance with the standard CEI en 60601-1-2 (Electro-medical apparatus, collateral standard:
electromagnetic compatibility) and therefore has a degree of immunity and emissions such as not to create dangerous interference with devices
complying with the same norm. Warning: Interference with electrical equipment that has a level of emission or immunity that does not conform to
CEI en 60601-1-2 may occur. In such cases, these equipment should not be used at the same time with O.M.S. equipment; If the device is in a
stalemate due to such interference, it is sufficient to turn it off and on again.
In the presence of particular holes in the supply voltage, it is possible that the micromotor has a momentary drop in speed that does not affect the
safety and performance of the same, the duration of the event is limited to the duration of the voltage drop.
The device is intended to work in the electromagnetic environment below specified. The customer or user of the device must ensure that it is used in this
environment.
Emission test
Emission RF
CISPR 11
Emission RF
CISPR 11
Harmonic emissions
IEC 61000-3-2
Fluctuations voltage emissions /flicker
IEC 61000-3-3
The device is intended to work in the electromagnetic environment below specified. The customer or user of the device must ensure that it is used in this
environment.
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Transients/rapid electrical pulse
sequence (Burst)
IEC 61000-4-4
Surge (Surge)
IEC 61000-4-5
Voltage gaps, short
interruptions and voltage
variations on the input lines
dell'alimentazione
Manufacturer's Guide and declaration – Electromagnetic emissions
Manufacturer's Guide and declaration – Electromagnetic immunity
Test level IEC 60601
contact ± 6 kV
air ± 8 kV
± 2 kV For power lines
± 1 kV For input/output lines
± 1 kV between the phases
± 2 kV between phases and ground
<5 % U
T
(>95 % hole in U
per 0.5 cycle)
T
OMS LINEA PATAVIUM S.T.01/3X ED.0 REV.8 09/2017
Complience
Electromagnetic Environment-Guide
Gruop 1
The device uses RF energy only for its internal operation. As a
result, its RF emissions are very low and probably does not cause
any interference in the electronic devices located nearby.
Class B
The device is suitable for use in all environments, including
domestic ones and those connected directly to a low-voltage
public network power supply that feeds buildings used for
domestic purposes.
Class A
In
complience
Level of
Compliance
± 6 kV
The floors must be in wood, concrete or ceramic tiles. If
the floors are covered with synthetic material, the relative
humidity should be at least 30%.
± 8 kV
± 2 kV
The quality of the main voltage should be that of a typical
commercial or hospital environment.
± 1 kV
± 1 kV
The quality of the mains voltage should be that of a typical
commercial or hospital environment
± 2 kV
<5 % U
The quality of the mains voltage should be that of a typical
T
commercial or hospital environment. If the user of the
device requires continuous operation during the
0.5 cycle (10mS)
interruption of the mains voltage, it is recommended to
power the device with a UPS or with batteries.
Electromagnetic Environment-Guide
65
Need help?
Do you have a question about the LINEA PATAVIUM and is the answer not in the manual?
Questions and answers