Fluid Detector
Assembly
P/N: SM-ACP215-01-EN(AB)
8. Thread the large knurled centrifuge knob (flat side up) onto the lock
shaft.
9. Install the o-ring into the channel of the centrifuge knob cap. Apply a light
coating of o-ring lubricant.
10.Apply a small amount of purple Loctite onto the top of the lock shaft and
thread the knob cap onto the shaft. Tighten the cap down using a small
set of needle nose or snap ring pliers.
11.Close centrifuge cover halves to check the general operation of the
assembly.
12.Apply a small amount of green Loctite to the underside edge of the knob
disk and set it onto the knob cap. Wipe off any excess Loctite.
13.Install the cover switch assembly and calibrate the switch position. (See
"Cover Switch Reassembly" on page
14.Install the centrifuge into the cabinet (see
page
142). Functionally check the operation of the centrifuge, cover
switch, and fluid detector.
"Fluid sensor test" on page
1. Disconnect the connector from the centrifuge distribution board.
2. Remove the (1) center Phillips head screw from the fluid detector
assembly pcb.
3. Gently twist the fluid detector assembly out of the centrifuge well.
143.)
"Centrifuge Reassembly" on
"Centrifuge cover and lock test" on page 8
8.
®
Haemonetics
ACP-215
Disassembly
and
®
Service Manual