Technical Data; Electrical Data - 3ctest EDS 20H User Manual

Hand-held electrostatic discharge simulator
Table of Contents

Advertisement

EDS 20H Hand-held Electrostatic Discharge

3 Technical Data

3.1 Electrical Data

3.1.1 EDS 20H General Data
Output
voltage
range
Voltage resolution
Polarity
Output
voltage
tolerance
Holding time
Default
Discharge
electrodes
Default RC module
RC
module
recognition
Controls
Repetition
frequency
Trigger modes
Impulse counter
Temperature
and
humidity
Interface
Battery duration
Battery type
Battery standby time > 30h
Battery
continuous
working time
Battery
charger
power requirements
Battery
charging
time
Web:
www.3ctest.cn
Simulator---User Manual
1.0kV to 18kV
100 V steps
Positive/negative, automatic switching
± 5%
> 5s
Round tip for air discharge / sharp tip for contact discharge according
to IEC 61000-4-2 & ISO 10605
150pF/330Ω, according to IEC/EN 61000-4-2 ed 2, waveform
specification see table 1 below
Automatic, values shown in display and test report
Touch screen, Trigger key, Smart key
Single,0.1,0.2,0.5,1,2,5,10 and 20 Hz
Auto &Manual
1 – 9999, pre-selectable
Temperature and humidity indicator inside
Fiber-optic
Contact discharge 18kV,1 Hz,approx.12h
Contact discharge
NiMH
12h(18KV,1Hz Contact discharge)
100V - 240V,47-63Hz,17VA
Approx.3.5h
3 /28
8kV,1 Hz,approx.18h
E-mail
3CTEST
20151021 V1.1
info@3ctest.cn