Thank you for choosing 3ctest to meet your ESD testing needs. Please take a little time to read through this user manual and familiarize yourself with the instrument controls and some potential dangers. We hope you have many productive years of operation from the EDS 20H Simulator.
3.5 General ............... 6 4. Initial Operation ............6 4.1 Visual Checks ............6 4.2 Installation ............6 5. EDS 20H Composition and Function ......... 7 5.1 Power ..............7 5.2 Parts ..............8 6. Menu Function ............12 6.1 Main menu ............12 6.2 Sub-menus ............
Never touch the equipment under test when EDS 20H is operating. If any part of the EDS 20H is damaged or it is possible that damage has occurred, for example during transportation, do not switch on the unit.
There are specific test requirements that must be satisfied when conducting ESD testing. These are noted in some detail in IEC/EN 61000-4-2, ISO 10605 and other standards. Because of its modular design, EDS 20H can be adapted quickly and easily to perform testing to any ESD standard.
Simulator---User Manual 3CTEST 20151021 V1.1 EDS 20H Hand-held Electrostatic Discharge 3 Technical Data 3.1 Electrical Data 3.1.1 EDS 20H General Data Output voltage 1.0kV to 18kV range Voltage resolution 100 V steps Polarity Positive/negative, automatic switching Output voltage ± 5%...
Simulator---User Manual 3CTEST 20151021 V1.1 EDS 20H Hand-held Electrostatic Discharge 3.2 Operating Conditions Operating Temperature Range +10~35° C Storage Temperature Range -10~+50° C Air humidity 20~80%r.h, non condensing 3.3 Mechanical Data Dimensions: approx. W 290mm x H 270mm x D 110mm Weight:...
Damaged packing cases may by a sign of transport damage. Damage caused in transit must be reported to the shipping agent immediately. If damage to EDS 20H or its charging unit can be seen or is suspected, do not apply any voltage! 4.2 Installation...
EDS 20H simulator comes together with two rechargeable NiMH batteries and a battery charger. The battery has to be either removed from EDS 20H to be charged, or charged while EDS 20H is not in operation. The battery charger cable has to be connected to the battery’s charging connector, located at the base (bottom) of the battery, and has to be correctly and securely inserted to ensure correct charging.
5.1.2 AC mains adapter (optional) The optional AC mains adapter set can be used for powering the EDS 20H from the power mains. To power the EDS 20H, replace the battery pack by the mains adapter and connect it to the power supply output cable.
Page 13
5.2.3 Tripod The EDS 20H connection position for the tripod is shown in figure 4. For continuous testing the use of a tripod is recommended. Screw thread is ¼ inch UNC. Fig.4 tripod position 5.2.4 Optical interface(optional)...
Page 14
EDS 20H is supplied with one contact discharge tip and one air discharge tip. Additional tips such as 30mm air discharge and fast rise time tips are available as an option.
Page 15
Additional modules are available. More details please refer to section ―Accessories and Options‖. To replace RC module, remove the cover at the side of EDS 20H as seen in figure 8. A small pull-out handle on the RC module is used for pulling out the RC module from EDS 20H.
20151021 V1.1 EDS 20H Hand-held Electrostatic Discharge Fig.8 RC module 6 Menu Function 6.1 Main menu EDS 20H is operated via a touch screen, the figure below resembles a screenshot of the main menu. Fig.9 EDS 20H main menu (1)...
Simulator---User Manual 3CTEST 20151021 V1.1 EDS 20H Hand-held Electrostatic Discharge (2) Set frequency (3) Power key (4) System setting (5) Discharge mode selection (6) RC network(display only) (7) Polarity set/display (8) Battery life display (9) Set voltage (10) Display running state 6.2 Sub-menus...
Page 18
Simulator---User Manual 3CTEST 20151021 V1.1 EDS 20H Hand-held Electrostatic Discharge (3) Left or right, move the cursor to select setting the bit of counter (4) Confirm (5) ▲▼ Up or down to set the counting value (6) Cancel 6.2.2 Frequency setting Fig.11 Frequency setting...
Page 19
Simulator---User Manual 3CTEST 20151021 V1.1 EDS 20H Hand-held Electrostatic Discharge Fig.12 System setting (1) Item list ▲▼ (2) Up/Down to select the item of system setting (3) Setting key (4) Back System setting items list: a) Backlight:on/off Set the backlight on/off of the screen b) Contrast:50% Touch Up/Down to set the screen contrast, the range is 10%~90%.
Page 20
Touch Up/Down to select trigger mode and set Auto/Manual. Note: Only can choose Manu for air discharge mode. e) Power Off Touch Up/Down to select Power off and set, EDS 20H will be powered off. 6.2.4 Voltage setting Fig.14 Voltage setting-1 (1)...
To find out more about test levels and test setups, please refer to the relevant standards. 2×470KΩ Accessories such as HCP, test table, resistor cable, etc. are available from 3ctest. Please refer to section ―Accessories and Options‖. Test setup example according to IEC/EN 61000-4-2 ed2.
Almost any major failure within one of the low voltage or the high voltage circuits would be detected and reported immediately. When errors occur during operation of EDS 20H, a warning message will appear. If an error occurs, please switch off the unit to avoid possible damage. Error messages are...
Page 23
Simulator---User Manual 3CTEST 20151021 V1.1 EDS 20H Hand-held Electrostatic Discharge ERROR 001 RC module missing ERROR 002 charging ERROR 003 Discharge tip missing 充电器 Temperature and humidity over ERROR 004 range Take measures for error message as following: Error Code...
Simulator---User Manual 3CTEST 20151021 V1.1 EDS 20H Hand-held Electrostatic Discharge 9 Accessories and Options 9.1 Accessories The following accessories are included with basic Version of EDS 20H: Carrying case User manual Air discharge tip Contact discharge tip ...
10.3 Calibration Calibration period of EDS 20H has to be determined by the user, and depends on the intensity of use and end user requirements. Typically, a calibration is recommended every 1 year.
Suzhou 3ctest, with "good faith", "innovation", "win-win" for the purpose of the company, to promote and applications as own duty, as "3ctest" flagship brand for the company, actively explores the domestic and foreign markets, extensively with colleges and universities, the national research institutions, measurement test department and the international company to take all-round cooperation.
Hong Kong and Taiwan and Southeast Asia. Suzhou 3ctest has always been to high-tech products as actively grasp of the future market competition, from product use to test, countermeasure and solution, we will, as always, actively...