3. Pre-treatment of work piece

To get the accurate measuring results, pre-treatment of work piece is
required.

3.1 Work piece surface

Temperature of work piece should be less than 80℃;
The surface roughness requirements are listed in table 3.1.1
Table: 3.1.1
Impact Device
D, DC, D+15, DL
G
C
The small support ring or non-conventional support rings are required for
work piece with curved surface radius less than 30mm, for details please refer
to appendix 4.

3.2 Weight of work piece

No support is required for work piece weight more than 5kg.
Work piece with weight of 2-5kg and also heavier work piece with
protruding parts or thin walls should be placed on a solid support. In such a
manner, they do not move or flex during testing.
Work piece less than 2 kg should be rigidly coupled with a non-yielding
support such as a heavy base plate.
Table 3.2.1 Work piece weight and thickness
Impact Device
D\DC\DL\D+15
G
C

3.3 Work piece surface hardened layer

If the work piece surface hardened layer is too thin, the impact force of
short duration will go through the layer and make the L-value incorrect. The
proper depth surface hardened layer is listed below:
Hardness Tester DHT-200
Work piece of surface roughness
Weight (KG)
Need
Need
coupled
support
0.1-2
2-5
0.5-5
5-15
0.2-0. 5
0.5-1.5
PLUS
2um
7um
0.4um
No
need
> 5
>15
>1.5
3
Minimum thickness
of work piece (mm)
3
10
1