11.2 Menu overview

Menu item
Adjustment mode
Reference quantity
Automatic
zero point correction
Baud rate
Data output
Sound by pressing
the button
Automatic internal adjustment
having connected the electrical
power (only ADJ models)
Filter
* = factory setting
26
Settings
Description
C1 - 0
Internal adjustment (ADJ models:
always use this setting)
C1 – 1
External adjustment (ADB models:
always use this setting)
C2 – 0*
10
C2 - 1
20
C2 - 2
50
C2 - 3
100
C2 - 4
1000
C3 - 0
No Zerot Tracking
C3 – 1*
1D Zero-point Tracking
C3 - 2
2D Zero-point Tracking
C3 - 3
3D Zero-point Tracking
C3 - 4
4D Zero-point Tracking
C3 - 5
5D Zero-point Tracking
C3 - 6
Not documented
C4 - 0
1200
C4 - 1*
2400
C4 - 2
4800
C4 - 3
9600
C5 - 0
Automatic output of stable weighing
values
C5 - 1
Via remote control command W (w)
C5 - 2
Continuous data output
C5 – 3*
Output for stable and instable weighing
values after pressing PRINT key
C6 – 0
switched on
C6 - 1*
switched off
C7 - 0*
switched on
C7 - 1
switched off
C8 - 0*
Low interference degree
C8 - 1
Medium interference degree
C8 - 2
High interference degree
C8 - 3
Not documented