Rigol DP700 Series Calibration Manual
Rigol DP700 Series Calibration Manual

Rigol DP700 Series Calibration Manual

Programmable linear dc power supply
Hide thumbs Also See for DP700 Series:

Advertisement

RIGOL
Calibration Guide
DP700 Series Programmable
Linear DC Power Supply
July 2016
RIGOL TECHNOLOGIES, INC.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the DP700 Series and is the answer not in the manual?

Questions and answers

Summary of Contents for Rigol DP700 Series

  • Page 1 RIGOL Calibration Guide DP700 Series Programmable Linear DC Power Supply July 2016 RIGOL TECHNOLOGIES, INC.
  • Page 3: Guaranty And Declaration

    Notices RIGOL products are covered by P.R.C. and foreign patents, issued and pending.  RIGOL reserves the right to modify or change parts of or all the specifications  and pricing policies at the company’s sole decision. Information in this publication replaces all previously released materials.
  • Page 4: Document Overview

    RIGOL Document Overview This manual gives a detailed description of the calibration notices and calibration process of DP700 series programmable linear DC power supply (hereinafter referred to as "DP700 series"). Main Topics in this Manual: Chapter 1 Calibration Notices This chapter gives a detailed description of the recommended calibration interval, calibration safety, how to stop the calibration, how to acquire the calibration service, the recommended calibration devices, and the precautions during calibration.
  • Page 5: Table Of Contents

    Contents RIGOL Contents Guaranty and Declaration ................. I Document Overview ................II Chapter 1 Calibration Notices ............. 1-1 Calibration Interval ................1-1 Calibration Safety .................. 1-1 To Stop the Calibration ................1-2 To Acquire the Calibration Service ............1-2 Recommended Calibration Devices ............1-2 Calibration Precautions ................
  • Page 7: Chapter 1 Calibration Notices

    RIGOL Chapter 1 Calibration Notices Calibration Interval A regular calibration can ensure the performance specifications of DP700 series. The recommended calibration interval is one year. Please perform a regular calibration if a higher accuracy is required for DP700 series. You are recommended to perform a complete calibration within the recommended calibration interval to guarantee its specifications before the next calibration.
  • Page 8: To Stop The Calibration

    To Acquire the Calibration Service RIGOL does not recommend users to perform manual calibration by themselves. If calibration is required, please contact RIGOL customer service department or the local distributor.
  • Page 9: Calibration Precautions

    Chapter 1 Calibration Notices RIGOL Calibration Precautions To achieve an optimal performance, you should follow the advices below to perform all the calibration procedures. Ensure that the ambient temperature stays between 18°C and 28°C. Ideally, the  calibration should be performed at 23°C±1°C.
  • Page 11: Chapter 2 Calibration Process

    Chapter 2 Calibration Process RIGOL Chapter 2 Calibration Process The predures for calibration process for DP700 series are as follows: Step 1 Enter the calibration page Step 2 Select the calibration item Step 3 Select the calibration point Step 4 Enter the measurement value...
  • Page 12 Step 2 Select the calibration item In the calibration page, press Up/Down arrow key or use the knob to switch the calibration items. The calibration items for DP700 series are available in Table 2-1. You can select one or multiple items for calibration. RIGOL recommends that you perform the calibration for all the available items.
  • Page 13 Chapter 2 Calibration Process RIGOL Note: During calibrating the items, ensure that no load is connected to the output  terminals. OVP calibration does not need any calibration devices. For its detailed  calibration operation, see the following contents. After you complete the calibration, restart the instrument.
  • Page 14 Figure 2-4 "Save the calibration data?" Dialog Box Note: If you want to calibrate multiple items, RIGOL suggests you complete the  calibration for all the items and then save the calibration data, so that you do not have to enter the calibration password repeatedly to enter the calibration page for calibrating each calibration item.
  • Page 15: Dac

    V DAC This section takes DP711 as an example to introduce the calibration method for the voltage DAC of DP700 series. Unless otherwise specified, this calibration method applies to all the models of the DP700 series. (1) Ensure that DP711 has been warmed up for at least 30 minutes. Use the test probe to connect the output terminals of DP711 to the voltage input terminals of the digital multimeter, as shown in Figure 2-5.
  • Page 16 RIGOL Chapter 2 Calibration Process (7) Select the rest of the calibration points according to the ID sequence, then repeat Step (5) and (6) until you complete the calibration operation for all the calibration points. (8) Press the Down arrow key or use the knob to switch the parameter focus to "Save", then, press OK to save the current calibration data.
  • Page 17: Adc

    V ADC This section takes DP711 as an example to introduce the calibration method for the voltage ADC of DP700 series. Unless otherwise specified, this calibration method applies to all the models of the DP700 series. (1) Ensure that DP711 has been warmed up for at least 30 minutes. Use the test probe to connect the output terminals of DP711 to the voltage input terminals of the digital multimeter, as shown in Figure 2-5.
  • Page 18: I Dac

    I DAC This section takes DP711 as an example to introduce the calibration method for the current DAC of DP700 series. Unless otherwise specified, this calibration method applies to all the models of the DP700 series. (1) Ensure that DP711 has been warmed up for at least 30 minutes. Use the test probe to connect the output terminals of DP711 to the current input terminals of the digital multimeter, as shown in Figure 2-6.
  • Page 19 Chapter 2 Calibration Process RIGOL (7) Select the rest of the calibration points according to the ID sequence, then repeat Step (5) and (6) until you complete the calibration operation for all the calibration points. (8) Press the Down arrow key or use the knob to switch the parameter focus to "Save", then, press OK to save the current calibration data.
  • Page 20: I Adc

    I ADC This section takes DP711 as an example to introduce the calibration method for the current ADC of DP700 series. Unless otherwise specified, this calibration method applies to all the models of the DP700 series. (1) Ensure that DP711 has been warmed up for at least 30 minutes. Use the test probe to connect the output terminals of DP711 to the current input terminals of the digital multimeter, as shown in Figure 2-6.
  • Page 21: Ovp

    RIGOL This section takes DP711 as an example to introduce the calibration method for the OVP of DP700 series. Unless otherwise specified, this calibration method applies to all the models of the DP700 series. (1) Ensure that DP711 has been warmed up for at least 30 minutes.