Product overview (Fig. 1)
1
On/off button
2
Status indicator
3
Detachable connector
4
Micro-USB socket of device
5
Battery indicator
6
Adapter
7
Socket for standard USB plug
8
Standard USB plug
9
Micro-USB plug
10 Self-adhesive hydrogel electrodes
Replacement
Electrodes
Replace the electrodes if:
-
they are damaged or torn.
-
they are past the use-by date indicated on the resealable bag.
-
they have lost their adhesive power. Never use plaster or tape to attach
them to your skin.
-
stimulation feels less strong.
-
when stimulation is uncomfortable, i.e. when you experience an
unpleasant stinging or biting sensation.
Note: Always replace the electrodes with electrodes recommended for this device
by the manufacturer. You can order new electrodes at www.philips.com/
tens-electrodes.
Guarantee and support
Your device has been designed and developed with the greatest possible
care to guarantee an expected service life of 5 years.
If you need service or information or if you have a problem, please visit our
website at www.philips.com/support. You can also contact the Philips
Consumer Care Centre in your country. You can reach this centre
at telephone number +971 6 5178727.

Guarantee restrictions

The terms of the guarantee do not cover the electrodes, as the electrodes
are subject to wear and have to be replaced regularly.
Manufacturer's legal address
Philips Consumer Lifestyle B.V.
Tussendiepen 4, 9206 AD Drachten, Netherlands
Recycling
Do not throw away the appliance with the normal household waste at
the end of its life, but hand it in at an official collection point for recycling.
By doing this, you help to preserve the environment.

Specifications

Model
Rated voltage(V) - adapter
Rated frequency (Hz) - adapter
Rated output - adapter
Class
Ingress of water - device
Battery type - device
Adapter
Pulse output parameters
Frequency range
Pulse duration
Maximum current output
Maximum output voltage
Current pulse shape
Operating conditions
Temperature
Relative humidity
Atmospheric pressure
PR3840
100-240 ~ 150mA
50-60
5V , 500 mA
Medical device Class IIa
IP22
Li-Ion
The adapter is specified as part of
the device.
1-100 Hz
60-350 µs for programs 1 to 20
60mA at 500-1000 Ohm
60V
Biphase symmetrical (net current 0 DC)
from +5°C to +40°C
from 15% to 93% (non-condensing)
from 700hPa to 1060hPa
Storage and transport conditions
Temperature - electrodes
Temperature - device
Relative humidity
Atmospheric pressure
Note: If you want to store the electrodes for more than a month, keep them at
temperatures between +5°C and +27°C.
Note: Do not store the electrodes in the freezer or refrigerator. Do not expose
them to high temperatures, nor immerse them in water, or leave them outside
of the plastic bag.
Explanation of symbols
Symbols on the adapter and the device
-
This symbol indicates that the adapter is double insulated (Class II),
according to IEC 60601-1 (Fig. 2).
-
This symbol means: Do not throw away with the normal household
waste. For further instructions, see chapter 'Recycling' (Fig. 3).
Symbols on the device
-
This is the symbol for serial number. This symbol is followed by the
manufacturer's serial number (Fig. 4).
-
This is the symbol for reference number. This symbol is followed by
the manufacturer's reference number, which is the type number of
the device (Fig. 5).
-
This symbol on the device means: protected against access to hazardous
parts with a finger and against vertically falling water drops when tilted
up to 15 degrees (Fig. 6).
-
This symbol means that the part of the device that comes into physical
contact with the user (also known as the applied part) is of type BF
(Body Floating) according to IEC 60601-1. The applied parts are the
electrodes (Fig. 7).
-
This symbol means that this device emits non-ionising radiation.
All devices and systems that include RF transmitters or that intentionally
apply RF electromagnetic energy must have a label with this symbol (Fig. 8).
-
This symbol means: caution TENS output. You find this symbol near all
electrode connections (Fig. 9).
-
This symbol means 'Manufactured by' and appears next to the address
of the legal manufacturer (Fig. 10).
-
This symbol means: Conforms to EC Directives. CE stands for 'Conformité
Européenne'. 0344 is the number of the notified body (Fig. 11).
-
This symbol means: read the user manual before you start using
the device (Fig. 12).
-
This symbol is the mark that identifies the standby button (Fig. 13).
-
This is the pulse symbol. It appears next to the status indicator (Fig. 14).
-
This is the battery symbol. It appears next to the battery indicator (Fig. 15).
Symbols on the adapter
-
This symbol means: alternating current (Fig. 16).
-
This symbol means: direct current (Fig. 17).
-
This symbol means: Conforms to EC Directives. CE stands for
'Conformité Européenne' (Fig. 18).
-
This symbol means: may only be used indoors (Fig. 19).
-
This symbol means: USB connector (Fig. 20).
-
This symbol means: micro-USB connector (Fig. 21).
-
This symbol is the certification logo that includes identification of the
accredited compliance test house. (Fig. 22)
-
This symbol is the certification mark for technical equipment.
The GS mark is based on the German Equipment and Product Safety
Act (Fig. 23).
-
This symbol shows that the no-load power CEC (California Energy
Commission) Efficiency Level is V (five) in order to meet EU
requirements (Fig. 24).
Symbols on electrode bag
-
This symbol means: Do not use while driving, operating machines or
performing other activities that may present a risk of injury (Fig. 25).
-
This symbol means: Do not use damaged or worn electrodes or
electrodes that have lost their adhesive power (Fig. 26).
-
This symbol means: The electrodes should be used by one person and
not shared with others (Fig. 27).
-
This symbol means: Do not place the electrodes on top of each other
or so close to each other that they touch each other (Fig. 28).
-
This symbol means: Do not place electrodes on your head or on
your neck (Fig. 29).
-
This symbol means: Do not place the electrodes on your chest (Fig. 30).
-
This symbol means: Do not put the electrodes on open wounds,
cancerous lesions, rashes and infected or inflamed skin (Fig. 31).
from 0°C to +40°C
from -10°C to +50°C
less than 93% (non-condensing)
from 700hPa to 1060hPa