Fuse Holder & Voltage Selector; Mains Cord; Connections - UGO BASILE 58500 Instruction Manual

Bp recorder
Table of Contents

Advertisement

UGO BASILE - I
4.3.2
Fuse Holder & Voltage Selector
The fuse holder comprises two fuses, one on the live, and the other on the neutral. For operation
at 230-240 Volts, we recommend 315 mA timed "slow blow" fuses (type T315). Use 630 mA
fuses (type T630) for operation at 115 Volts.
The fuse holder also embodies the Voltage Selector. Make sure that the label visible in the win-
dow indicates the correct voltage (i.e., the voltage of your mains).
To replace the fuses and/or change the selected voltage, see paragraph 12.1-Electrical.
4.3.3

Mains Cord

It is a standard cable, Cat. # E-WP008. Make sure your power outtake is provided with a reliable
ground connection, see also 4.4 & 4.5.
4.4

Connections

Connect the mains cord to a power outtake, provided with a reliable earth connection.
Beside the power module described in paragraph 4.3, the BP Recorder back panel encompasses
the following connectors, from left to right:-
SERIAL INTERFACE:
PRESSURE SIGNAL
(ANALOGIC):
PULSE SIGNAL:
TALY
this delta 9-pin connector enables the operator to connect the in-
strument to the serial port RS232C of the PC, see paragraph 3-
DATA ACQUISITION.
these connectors are useful to route both pressure and pulse
signals to CRT (oscilloscope), chart recorder or PC (via a data
acquisition system)
Series 58000 Instruction Manual (Rev.1)
Figure 10 "Back Panel"
Page 10