UGO BASILE - I
In either case, the display shows the following picture,
which illustrates the diastolic pressure pattern (if previously
set).
The cursor mark is over the graph and the pressure figure
is below the base line.
As in the previous case, the cursor can be shifted to select
the best possible position.
In the case the Chart Recorder is on, the final picture of
the experiment is recorded (see the example taken from
a real experiment) where we read, from the top of the
data set:
⇒
date
⇒
time
⇒
systolic pressure (compression run)
⇒
diastolic pressure (compression run)
⇒
systolic pressure (decompression run)
⇒
diastolic pressure (decompression run)
⇒
pulse rate
⇒
gender
⇒
animal number
The latter two data, gender and animal number, are represented in the display as, e.g.,
M1 which means animal No. 1, male, F3 which means animal No. 3, female, etc.
CALIBRATION
8
To calibrate means to compare the pressure reading of 58500 with the data supplied by a
precision mercury pressure gauge (or any other certified pressure standard) capable to
cover the range 50 – 300 mm Hg.
TALY
The datum (in our case the systolic pressure) can be
saved
(NUL) key as shown in the display at this side.
by the F4 (CON) key or cancelled by the F3
2003/02/05
15:25:13
Systolic Up
Diastolic Up
Systolic Down
Diastolic Down
Pulse
Gender Male
Animal N° 12
Series 58000 Instruction Manual (Rev.1)
121
070
116
072
Rate
545
Page 20