Bio-Optica BEC150 Instructions For Use Manual

Bio-Optica BEC150 Instructions For Use Manual

Paraffin embedding station

Advertisement

Instructions for use
BEC150
Paraffin embedding station
In vitro diagnostic - medical device
CND code: W0202059002
Bio-Optica code: 40-200-002
Reference number code: 1831667
Always keep this manual near the instrument.
Read thoroughly before operating this instrument.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the BEC150 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Bio-Optica BEC150

  • Page 1 Instructions for use BEC150 Paraffin embedding station In vitro diagnostic - medical device CND code: W0202059002 Bio-Optica code: 40-200-002 Reference number code: 1831667 Always keep this manual near the instrument. Read thoroughly before operating this instrument.
  • Page 2 User Manual Publication date: 02/03/2020 Rev. 00 Bio-Optica Milano S.p.A. Via San Faustino 58 - 20134 Milan Phone +39 02.21.27.13.1 - Fax Italy +39 02.21.53.000 - Fax Export +39 02.21.54.155 www.bio-optica.it – info@bio-optica.it – claims@bio-optica.it...
  • Page 3 Telephone (+39) 02-2127131 Fax: (+39) 02-2153000 Internet: info@bio-optica.it Publication date: 02/03/2020 Rev. 00 Bio-Optica Milano S.p.A. Via San Faustino 58 - 20134 Milan Phone +39 02.21.27.13.1 - Fax Italy +39 02.21.53.000 - Fax Export +39 02.21.54.155 www.bio-optica.it – info@bio-optica.it –...
  • Page 4: Table Of Contents

    CLEANING AND MAINTENANCE ..........................25 DECLARATION OF CONFORMITY ..........................26 Publication date: 02/03/2020 Rev. 00 Bio-Optica Milano S.p.A. Via San Faustino 58 - 20134 Milan Phone +39 02.21.27.13.1 - Fax Italy +39 02.21.53.000 - Fax Export +39 02.21.54.155 www.bio-optica.it – info@bio-optica.it –...
  • Page 5: Important Information

    They provide important information regarding the functions and safety in installing, using and maintaining the instrument. Bio-Optica Milano S.p.A. cannot be held liable for any damage resulting from improper or incorrect use and for failure to comply with the provisions of this manual and the current safety standards.
  • Page 6: Symbols And Their Meaning

    WARNING symbol: ATTENTION or WARNING symbol: hot surface temperatures Publication date: 02/03/2020 Rev. 00 Bio-Optica Milano S.p.A. Via San Faustino 58 - 20134 Milan Phone +39 02.21.27.13.1 - Fax Italy +39 02.21.53.000 - Fax Export +39 02.21.54.155 www.bio-optica.it – info@bio-optica.it –...
  • Page 7: Proper Use

    The BEC150 is a modern embedding station for a practical, quick and accurate embedding of paraffin samples. Thanks to the integration of the thermal unit, the BEC150 can find a place in any pathological anatomy laboratory. The main functions of the unit are: •...
  • Page 8: Safety Warnings

    Use the Bio-Optica spatula supplied together with the instrument (40-200-070). During operation, the paraffin tank, the container for the molds, the tray for the cassettes, the workspace and the forceps-holder are at high temperatures. Risk of burns! Do not store any combustible or flammable substance near the instrument.
  • Page 9: Description

    5. Dispenser with lever and lighting 6. Workspace 7. Heated forceps container Publication date: 02/03/2020 Rev. 00 Bio-Optica Milano S.p.A. Via San Faustino 58 - 20134 Milan Phone +39 02.21.27.13.1 - Fax Italy +39 02.21.53.000 - Fax Export +39 02.21.54.155 www.bio-optica.it – info@bio-optica.it –...
  • Page 10: Main Characteristics

    Aluminum work plate with no. 6 housings for forceps and paraffin collection cassette, both heated. Publication date: 02/03/2020 Rev. 00 Bio-Optica Milano S.p.A. Via San Faustino 58 - 20134 Milan Phone +39 02.21.27.13.1 - Fax Italy +39 02.21.53.000 - Fax Export +39 02.21.54.155 www.bio-optica.it –...
  • Page 11 Programmability: Turning the instrument on and off, weekly. Publication date: 02/03/2020 Rev. 00 Bio-Optica Milano S.p.A. Via San Faustino 58 - 20134 Milan Phone +39 02.21.27.13.1 - Fax Italy +39 02.21.53.000 - Fax Export +39 02.21.54.155 www.bio-optica.it – info@bio-optica.it –...
  • Page 12: Installation

    Do not use extension cables or adapters and do not modify the cable supplied. Publication date: 02/03/2020 Rev. 00 Bio-Optica Milano S.p.A. Via San Faustino 58 - 20134 Milan Phone +39 02.21.27.13.1 - Fax Italy +39 02.21.53.000 - Fax Export +39 02.21.54.155 www.bio-optica.it –...
  • Page 13: Use

    A built-in (removable) filter (9) Publication date: 02/03/2020 Rev. 00 Bio-Optica Milano S.p.A. Via San Faustino 58 - 20134 Milan Phone +39 02.21.27.13.1 - Fax Italy +39 02.21.53.000 - Fax Export +39 02.21.54.155 www.bio-optica.it –...
  • Page 14 (7), one for low level and one for high level for the paraffin. When the sensors are not covered by paraffin, a warning is issued to remind you to refill the tank with this visual message: Recycled paraffins cannot be used in the BEC150 due to the risk of contamination. Dispenser (2) The dispenser is heated separately.
  • Page 15 Jack for power supply to heated forceps and pestles (11) Under the jack logo, there are two inputs to connect the heated Bio-Optica forceps directly to the machine. With the two inputs it is possible to facilitate work for ambidextrous operators or to work with forceps and pestle at the same time.
  • Page 16: Operation

    4 – SETTINGS MENU: Enables access to functions (see Settings menu) Publication date: 02/03/2020 Rev. 00 Bio-Optica Milano S.p.A. Via San Faustino 58 - 20134 Milan Phone +39 02.21.27.13.1 - Fax Italy +39 02.21.53.000 - Fax Export +39 02.21.54.155 www.bio-optica.it –...
  • Page 17 The machine can operate both in MANUAL mode (par. 5.5) and in AUTOMATIC mode (par. 5.4). The BEC150 offers the option of using an integrated circuit to power the electrically-heated forceps and pestles. Once the FORCEPS mode is activated, the image of the machine will show two forceps-status check LEDs (as shown in the image).
  • Page 18: Settings Menu

    2 – Weekly work schedule 3 – Adjustment of operating temperatures Publication date: 02/03/2020 Rev. 00 Bio-Optica Milano S.p.A. Via San Faustino 58 - 20134 Milan Phone +39 02.21.27.13.1 - Fax Italy +39 02.21.53.000 - Fax Export +39 02.21.54.155 www.bio-optica.it –...
  • Page 19: Time And Date

    Publication date: 02/03/2020 Rev. 00 Bio-Optica Milano S.p.A. Via San Faustino 58 - 20134 Milan Phone +39 02.21.27.13.1 - Fax Italy +39 02.21.53.000 - Fax Export +39 02.21.54.155 www.bio-optica.it –...
  • Page 20: Manual Mode

    The display will now show the temperatures of the various parts; these will be adjustable from the appropriate control panel (par. 5.6). Publication date: 02/03/2020 Rev. 00 Bio-Optica Milano S.p.A. Via San Faustino 58 - 20134 Milan Phone +39 02.21.27.13.1 - Fax Italy +39 02.21.53.000 - Fax Export +39 02.21.54.155 www.bio-optica.it –...
  • Page 21 Warning: the manual operating mode can be used to extend a programmed work session at the end of the session itself Publication date: 02/03/2020 Rev. 00 Bio-Optica Milano S.p.A. Via San Faustino 58 - 20134 Milan Phone +39 02.21.27.13.1 - Fax Italy +39 02.21.53.000 - Fax Export +39 02.21.54.155 www.bio-optica.it –...
  • Page 22: Temperature Settings

    Shows useful information for contacting the technical assistance service or for reporting anything to the company. Publication date: 02/03/2020 Rev. 00 Bio-Optica Milano S.p.A. Via San Faustino 58 - 20134 Milan Phone +39 02.21.27.13.1 - Fax Italy +39 02.21.53.000 - Fax Export +39 02.21.54.155 www.bio-optica.it –...
  • Page 23: Operator Panel

    3 – Language (Italian/English) 4 – Enable/Disable Paraffin refill alert Publication date: 02/03/2020 Rev. 00 Bio-Optica Milano S.p.A. Via San Faustino 58 - 20134 Milan Phone +39 02.21.27.13.1 - Fax Italy +39 02.21.53.000 - Fax Export +39 02.21.54.155 www.bio-optica.it – info@bio-optica.it –...
  • Page 24 User Manual Publication date: 02/03/2020 Rev. 00 Bio-Optica Milano S.p.A. Via San Faustino 58 - 20134 Milan Phone +39 02.21.27.13.1 - Fax Italy +39 02.21.53.000 - Fax Export +39 02.21.54.155 www.bio-optica.it – info@bio-optica.it – claims@bio-optica.it...
  • Page 25: Cleaning And Maintenance

    The instrument does not need to be sterilized as the handling of fresh samples is not expected, only of histological samples fixed and embedded in paraffin. Maintenance Periodic annual maintenance must be carried out by qualified personnel authorized by Bio-Optica. For information, please contact the Technical Assistance Service at Tel. 02-21271310. Fuse replacement If the instrument does not switch on, check that the power cable has been connected properly, that there is current in the mains, and that the two fuses below the instrument’s power socket are not burnt.
  • Page 26: Declaration Of Conformity

    Nom, fonction CARLO SBONA Nombre, función Legal Representative Publication date: 02/03/2020 Rev. 00 Bio-Optica Milano S.p.A. Via San Faustino 58 - 20134 Milan Phone +39 02.21.27.13.1 - Fax Italy +39 02.21.53.000 - Fax Export +39 02.21.54.155 www.bio-optica.it – info@bio-optica.it –...