Your Maxi Sky 440 is part of a Maxi Sky 440. It will help protect your product as well as ensure series of quality products designed that the equipment performs to your satisfaction.
2 years from purchase date. The life expectancy only applies if the slings The Maxi Sky 440 is designed to assist caregivers in and stretchers have been cleaned, maintained and hospitals, long-term care, nursing homes and home inspected in accordance with the “Preventive...
General Information Symbols Used General Key to symbols Symbols General Key to symbols Symbols Alternating current. Points out the date of manufacture and the address of the manufacturer. Points out a type BF applied CE marking indicating part. conformity with European Community harmonised Indicates the product is a legislation.
Addendum for Rail Systems other than KWIKtrak™ Arjo could have adapted this lift cassette to be used with a rail system other than KWIKtrak. If this is the case, an addendum (001-14350-XX) must be supplied with this Instructions for Use. This document provides important information relative to: •...
100%. • Do not incinerate the battery. • The ambient temperature range when storing the Maxi Sky 440 remains between –25ºC and 70ºC (-13ºF to 158ºF). WARNING: This product contains small parts that might present a choking hazard for children if swallowed or inhaled.
Practices NOTE: Rigorous cleaning actions should be WARNING: Following the instructions is done when the Maxi Sky 440 is exposed to important for the safe use of the battery and an animal. Pet hair trapped inside the device to keep the user (resident/caregiver) from (through the strap opening) can reduce the harm.
Product Description Hand Control Charger The Maxi Sky 440’s hand control unit is used to The Maxi Sky 440 units are equipped with a wall- operate the portable lift. Each function is adapted charger (see Fig. 3) that can be described in .
Reacher 24'' & 36'' To use the Reacher, refer to the Reacher IFU. Fig. 6 Refer to your local Arjo agent for ordering. WARNING: All trolleys are not compatible with all rail systems. Please contact your local Arjo agent to get more information.
Make sure to have on hand a sling that is of the correct model and of adequate size for transfer with the Maxi Sky 440. Before performing the transfer, it is important to evaluate the patient’s general condition. For example, an agitated patient will require a more specific type of sling.
How to Use the Maxi Sky 440 Method 1 - Cross-through (Legs closed with crossing straps) WARNING: This method might not be suitable for patients with no upper body This method is recommended for most general control as they can slide down and almost transfer (see Fig.
How to Use the Maxi Sky 440 12) When the Maxi Sky 440 is no longer required, plug the portable lift back to the charger. WARNING: Before lifting the patient: 1) Make sure that all straps are attached to WARNING: Constant attention to the patient the lift’s supports.
How to Use the Maxi Sky 440 Fig. 14 Emergency Brake WARNING: Never use a portable lift with the emergency brake deployed. Use other The emergency brake is an additional safety equipment to lower the patient. Forcing the feature that automatically prevents the patient lift to make it move might lead to patient fall.
In order to reduce the charging time, refrain one in stock. from completely draining the batteries and leave The batteries included in the Maxi Sky 440 the portable lift connected to the charger between provide 10 to 30 transfers per battery charge.
WARNING: Safety related maintenance and authorized service must be carried out by qualified personnel, fully trained in servicing procedures by Arjo, and equipped with correct tools and proper documentation, including Parts List and Service Manual. Failure to meet these requirements could result in injuries and/or unsafe equipment.
Care and Maintenance Inspections by an Authorized Service Technician Before Every 2 Every 4 Every Every 2 Action/check Initially every use months months year years Replace strap. Inspect the welding aspect on the frame to detect cracks. Inspect transmission. Inspect connecting joints for proper attachment.
(see ). If there is any evidence of damage, DO NOT USE 3) Rinse with clean water any remaining residual . Contact your local Arjo agent to have the lift detergent. Then wipe clean with a dry cloth. serviced. •...
2 weeks. This will prevent premature other. aging of batteries. Strap Inspection If you store or ship the Maxi Sky 440, ensure that the power is turned off (emergency button pushed If the strap is damaged or shows signs of wear or in) beforehand.
Arjo agent assistance. Sling Inspection and Care See the documentation that comes with the sling. Annual Inspection The Maxi Sky 440 must be inspected annually by a qualified technician. WARNING: Maxi Sky 440 accessories must serviced every 12 months as a minimum requirement.
Troubleshooting Troubleshooting WARNING: Do not attempt to open the Maxi Sky 440 portable lift cassette. Only a qualified technician is authorized to open it. Alterations made to the portable lift by someone other than a qualified technician may cause serious injury.
WARNING: Portable RF communications equipment (including peripherals such as antenna cables & external antennas) should be used no closer than 30 cm to any part of the Maxi Sky 440, including cables specified by the manufacturer. Otherwise, performance degradation of this equipment could result.
Technical Specifications Seal lead-acid, rechargeable, recyclable. All batteries in the product must be Batteries recycled separately. Batteries are to be disposed in accordance with national or local regulations. Separate and recycled. Components that are primarily be made up of different The lift kinds of metal (containing more than 90% metal by weight) for example trolleys, sling bars, rails, upright supports, etc., should be recycled as metals.
Electromagnetic Emissions - For all Equipment and Systems The Maxi Sky 440 is intended for use in the electromagnetic environment indicated below. The customer or the user of the Maxi Sky 440 should assure that it is used in such an environment. Emissions test...
Electromagnetic Immunity - For all Equipment and Systems The Maxi Sky 440 is intended for use in electromagnetic environment specified below. The customer or the user of the Maxi Sky 440 should assure that it is used in such an environment. Electromagnetic...
Page 31
Electromagnetic Compatibility (continued) Guidance and Manufacturer’s Declaration - Electromagnetic Immunity - For all Equipment and Systems Electromagnetic Immunity test IEC 60601 test level Compliance level environment - guidance 3 V outside ISM bands 3 V outside ISM bands between 0.15-80 MHz Conducted RF between 0.15-80 MHz IEC 61000-4-6...
Page 35
AUSTRALIA FRANCE POLSKA Arjo Australia Arjo SAS Arjo Polska Sp. z o.o. Building B, Level 3 2 Avenue Alcide de Gasperi ul. Ks Piotra Wawrzyniaka 2 11 Talavera Road CS 70133 PL-62-052 KOMORNIKI (Pozna ) Macquarie Park, NSW, 2113, Australia...
Page 36
At Arjo, we are committed to improving the everyday lives of people affected by reduced mobility and age-related health challenges. With products and solutions that ensure ergonomic patient handling, personal hygiene, disinfection, diagnostics, and the effective prevention of pressure ulcers and venous thromboembolism, we help professionals across care environments to continually raise...