Replace the battery as follows, refer to Figure 2:
1. Turn the rotary switch to OFF and remove the test leads from the terminals.
2. Remove the six Phillips-head screws from the case bottom and remove the
battery door ().
While lifting the battery door, ensure the rubber gasket stays
attached to the battery compartment barrier.
3. Remove the three batteries and replace all three with AA Alkaline batteries
().
4. Ensure the battery compartment gasket () is properly installed around the
outside edge of the battery compartment barrier.
5. Replace the battery door by aligning the battery compartment barrier with
battery compartment.
6. Secure the door with the six Phillips-head screws.
How to Replace the Fuses
Referring to Figure 2, examine or replace the Meter's fuses as follows:
1. Turn the rotary switch to OFF and remove the test leads from the terminals.
2. Refer to step 2 under the How to Replace the Batteries section above to
remove the battery door.
3. Remove the fuse compartment seal () from the fuse compartment.
4. Gently lift out the fuse compartment door () from the fuse compartment.
5. Remove the fuse by gently prying one end loose, then sliding the fuse out of
its bracket ().
6. Install ONLY specified replacement fuses with the amperage, voltage, and
speed ratings shown in Table 5. The 440-mA fuse is shorter than the 10-A
fuse. For correct placement of each fuse, note the marking on the printed
circuit board under each fuse.
Note
Digital Multimeter
Basic Maintenance
9