Incompatible parts or accessories can result in degraded
performance�
Repairs and adjustments must be performed by Philips Respironics
authorized service personnel only� Unauthorized service could cause
injury, invalidate the warranty, or result in costly damage�
Periodically inspect electrical cords, cables, and the power supply for
damage or signs of wear� Discontinue use and replace if damaged�
To avoid electric shock, unplug the device and remove the batteries
before cleaning the bag� DO NOT immerse the device in any fluids�
For proper operation, your device requires unobstructed ventilation.
Always make sure any openings in the case are not obstructed by
items which may impede ventilation� Do not place the device in a
small closed space (such as a closet)� The device should not be used
adjacent to or stacked with other equipment� For more information,
contact your equipment provider�
Do not use an extension cord.
Device operation above or outside of the voltage, breath rate,
temperature, humidity and/or altitude values specified may decrease
oxygen concentration levels�
Be aware that the electrical cord and/or tubing could present a
tripping or strangulation hazard�
Use only power cords supplied by Philips Respironics for this device.
Use of power cords not supplied by Philips Respironics may cause
overheating or damage to the device and may result in increased
emissions or decreased immunity of the equipment or system�
Do not operate without the battery installed and working. If primary
power is lost with no battery in place, the device will stop operating
without warning the user� If the device must be operated without the
battery, the user should be aware that there is no backup power and
no warning if primary power is lost�
If the SimplyGo Mini device will not be used for a long period of time,
remove the battery from the device�
Place the device in a location so as to avoid pollutants and fumes.
Do not modify this system or equipment in any way. Modifications
could result in hazards to the user�