EMC Information

Your unit has been designed to meet EMC standards throughout its Service Life
without additional maintenance� There is always an opportunity to relocate your
portable SimplyGo Mini within an environment that contains other devices with
their own unknown EMC behavior� If you believe your unit is affected by locating
it closer to another device, simply separate the devices to remove the condition�
Portable Oxygen Concentrator
The SimplyGo Mini is designed to provide enriched oxygen� If there is a condition
where the device is not providing enriched oxygen, the device will alarm�
Oxygen Conserving
The SimplyGo Mini is designed to provide oxygen in response to the patient's
breathing� The device includes a sensing circuit that will detect when the patient
has started to inhale� If the SimplyGo Mini is either randomly or not providing
pulse dosed oxygen, this is a sign that the detection sensor is currently having
trouble detecting the patient's breathing pattern�
If there are errors or alarms affecting the purity or conserving features of the
device, the user should attempt to move the SimplyGo Mini to a different area to
determine if the issue is due to electromagnetic interference with other equip-
ment in the vicinity�
Guidance and Manufacturer's Declaration – Electromagnetic Emissions
SimplyGo Mini is intended for use in the electromagnetic environment specified below� The user of SimplyGo Mini should
assure that it is used in such an environment�
Emissions Test
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic Emissions
IEC 61000-3-2
Voltage Fluctuations/Flicker Emissions
IEC 61000-3-3
RF Emissions
CISPR25
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
SimplyGo Mini is intended for use in the electromagnetic environment specified below� The user of SimplyGo Mini should
assure that it is used in such an environment�
Immunity Test
IEC 60601 Test Level
Electrostatic
±8 kV Contact
Discharge (ESD)
±15 kV Air
IEC 61000-4-2
AMSR
46
Compliance
Group 1
Class B
Class A
Complies
Compliance Level
±8 kV Contact
±15 kV Air
oxygenconcentratorstore.com
Electromagnetic Environment - Guidance
The Device uses RF energy only for its internal
function� Therefore, its RF emissions are very
low and are not likely to cause any interference
in nearby electronic equipment�
The Device is suitable for use in all
establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network
that supplies buildings used for domestic
purposes�
The Device is suitable for use in vehicles and
boats�
Electromagnetic Environment - Guidance
Floors should be wood, concrete or ceramic tile�
If floors are covered with synthetic material, the
relative humidity should be at least 30%�