Page 1
Operation Manual M2000 Series – P/N 160988-10, Version 0.0 M2000 Series Power Analyzers ADAPTIVE Power Systems Worldwide Supplier of Power Equipment Copyright 2019, Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS.
Page 2
PAGE LEFT INTENTIONALLY LEFT BLANK FOR HARDCOPY VERSIONS OF THIS DOCUMENT 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 3
Harmonic Watts Measurements ......................45 4.7.1 Examples ................................45 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 3 of 284...
Page 4
6.7.2 Setting Power and Measurement User Preferences ..................74 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 4 of 284...
Page 5
Application Specific Measurements..................143 Overview .............................. 143 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 5 of 284...
Page 6
Comma Separated Value .CSV Files ....................... 213 9.6.4 Measurements File ............................213 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 6 of 284...
Page 7
11 Calibration ..........................274 11.1 Overview .............................. 274 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 7 of 284...
Page 8
12 CE MARK Declaration of Conformity..................280 Index ............................. 281 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 8 of 284...
Page 9
Figure 6-4: POWER DATA Screen .......................... 94 Figure 9-1: USB Drive Location and Info Field ....................208 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 10
Phone: +86-21-6763-9223 Fax: +86-21-5763-8240 Email: support@adaptivepower.com Web: http://www.adaptivepower.com 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 10 of 284...
Page 11
2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 12
Caution: Refer to this manual before using this product. Caution, risk of electric shock 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 13
Verify that the product is set to match the available line voltage and the correct fuse is installed. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 14
2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 15
4, “Technical Specifications”. Figure 3-1: Front and Rear Views 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 15 of 284...
Page 16
Low-pass filtering according to the BANDWIDTH configuration setting. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 16 of 284...
Page 17
For example, neither the measurement results nor the calibration scaling applied are modified for signal frequency content. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 18
For use with an external shunt or low impedance voltage output CT (20μV to 15Vrms) with resolution down to 0.1μV. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 19
Voltage and current DC (the average of the samples in the measurement period). 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 20
100 readings per second with no reduction in accuracy (up to 500/sec with reduced accuracy with W type channels). 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 21
- 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 22
Safety - CE mark certified to EN61010 with certified conformance to CAT I, II, III and IV mains measurement applications 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 23
N x 1ø There are no VPA measurements performed. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 23 of 284...
Page 24
Square root of the average squared sample in the measurement period. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 25
Each channel individually configurable for PT and/or CT use 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 25 of 284...
Page 26
The two VPAs may be configured as either – 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 26 of 284...
Page 27
Each VPA can be configured as N x 1ø, 2ø3w, 3ø3w(2ch), 3ø3w(3ch) or 3ø4w limited by the total number of channels installed. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 28
- or a partial measurement period time - if integration was started and/or stopped during it. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 29
Refer to section 10, “Remote Control Programming” on page 232 for details on using remote control. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 30
(e.g. if the total 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 31
Impedance (DC, AC and AC+DC), AC series resistance and inductance, AC parallel resistance and capacitance 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 32
Figure 4-1: Voltage Measurement base accuracy chart for AC MAINS applications 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 33
Figure 4-3: Voltage Measurement base accuracy chart for 50 kHz applications Figure 4-4: Voltage Measurement base accuracy chart for DC applications 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 34
Adjacent Channel Error 300 nV per V x Hz 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 34 of 284...
Page 35
Common-Mode = if 3ø3w(2ch) wiring = 1μV*(230V/√3)*50Hz = 6.6mV at (30+90)° = ±3.3mV 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 36
50mV/230V = 0 to +0.00022 (i.e. biased to a higher reading) 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 37
Figure 4-5: AD Card Current Measurement base accuracy chart for AC MAINS applications 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 38
Figure 4-8: AD Card Current Measurement base accuracy chart for DC applications 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 39
Figure 4-11: AX Card Current Measurement base accuracy chart for AC 50kHz applications 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 40
Figure 4-12: AX Card Current Measurement base accuracy chart for DC applications 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 41
7nV per V*Hz 0.2nV per V*Hz Error V*Hz 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 41 of 284...
Page 42
Bandwidth Limit Scaling = none when using AUTO-TRACK Total Scaling errors = ± 0.035% of reading 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 43
5mA / 5A = 0 to +0.001 (i.e. biased to a higher reading) 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 44
Alternately, as a worst case (at PF=1) this can expressed as F*0.013% of VA 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 45
Base Floor = (240 * 38μA ) + (3 * 45μV ) = 9255μW = ±9.255mW DC floor = 0 (signals have insignificant DC content) 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 46
(1mV / V AC+DC AC+DC AC+DC AC+DC 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 46 of 284...
Page 47
(1 – cos ( cos-1(1) + 0.015° * 0.05 )) = (1 – cos (0° + 0.015° * 0.05 )) = 0 Total Phase Errors = 0 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 48
2 frequency measurement periods 4 cycles of the signal 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 48 of 284...
Page 49
LF/PERIOD measurement period • Harmonic Measurement Period (from above) 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 49 of 284...
Page 50
As Voltage and Current Specifications for PK data (Watts = multiplication of V and A waveforms) 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 51
Zero to 99 days, 99 hours, 99 minutes, 99 seconds (1 second resolution) 0.01% + 8ms maximum error 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 52
9600, 19200, 57600, 115200 bps Formats Data, Parity, Stop Bits: 8,N,1 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 52 of 284...
Page 53
Max. Operating Altitude 2000 meters / 6562 feet ASL 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 53 of 284...
Page 54
If any of these items is missing, contact Adaptive Power Systems support department. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 55
A PEN OR A SCREWDRIVER OR A SIMILAR IMPLEMENT WITH SHARP CORNERS AS IT MAY DAMAGE THE TOUCH PANEL. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 56
Adaptive Power Systems or its authorized representative to arrange for service. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 57
2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 58
(both a scale factor in V/A and an offset are available). 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 59
ADJUST setting in the MEAS CONFIG screen). 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 60
2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 61
(both a scale factor in V/A and an offset are available). 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 62
In the MEAS CONFIG menu, set SCALE for the channels to OFF. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 63
(both a scale factor in V/A and an offset are available). 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 64
The rack height required in 4U or 7” / 177 mm. Figure 5-4: M2000-RMS Rack Mount Kit accessory 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 65
In some cases it shows an associated entered value 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 66
The present setting is shown in each button in a smaller text size, if you see anything that needs changing, press the corresponding button to change it. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 67
DIGITS: Set to 5. d. ADJUST: Set to NONE. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 67 of 284...
Page 68
If you turn off the unit. When you turn it back on it is 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright...
Page 69
Interface Time & Date Integration Measurement Info Info Status Status 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 69 of 284...
Page 70
POSSIBLE AT THIS TIME not be executed at the time received. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 71
Note: If operating the power analyzer as a unit in a multi-unit group then this holds or releases measurements in all power analyzers within that group. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 72
For the LAN interface only: 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 72 of 284...
Page 73
Main Firmware version FPGA Version 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 73 of 284...
Page 74
Using this screen, you can set your preference for turning on/off the power analyzer and your overall measurement preferences. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 75
The AUTOZERO button allows you to select if the power analyzer occasionally checks if the DC zero requires adjusting for changes in the operating environment. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 76
Total VA for each VPA depends on the WIRING selection for each VPA: 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 77
The power analyzer supports several methods of showing phase results as 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 78
Phase results in the range 0° to +180° are unaffected; results in the 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 79
VECTORS screen, no other data or results are affected by this selection. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 80
LAN interface can be used to control the unit from a computer. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 81
2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 82
The type of upgrade is accomplished by pressing the OPTION UPGRADE button and entering the factory supplied code. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 83
VPAs. If mechanical measurements are being performed, those will also be similarly 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 84
Pressing the ALL button selects that all available channels are used in the selected VPA. Note: Only channels that are installed are displayed. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 85
VPA), the channel scaling and offset settings stay with the channel. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 86
VPA if MEAS MODE is not EN61000-3-2 or EN61000-3-12. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 87
Fundamental frequency used for this VPA is automatically obtained from that VPA in real time. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 88
2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 89
The frequency resolution entry is limited to be within 0.01Hz to 1kHz 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 90
2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 91
The WIRING button allows you to select how measurements are combined and labelled when producing total results for this VPA. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 92
Press the CONFIGURATION button. The right area of the screen shows a list of the available configuration files. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 93
7. The file is then written. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 93 of 284...
Page 94
This screen shows the Volts, Amps, Watts, VA, VAR, PF and Frequency measurements for the selected Channel. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 95
‘clipped’ signals. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 96
2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 97
Inrush measurements. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 98
IMPEDANCE data is always shown (and is the same data in either screen). 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 99
See section 15 for full details. Integration results are always cleared when integration is started. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 100
The primary Volts and Amps data is the average for all phases/channels in the VPA. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 101
2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 102
See section 8.3 for full details regarding using this screen to make Inrush measurements. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 103
See section 8.4 for full details. Integration results are always cleared when integration is started. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 104
1. Insert the drive into the front panel USB port and wait for it to be READY. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright...
Page 105
2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 106
The displayed data is controlled by the buttons across the top of the screen (in order from left to right): 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 107
The numerical listing down the left side shows the amplitude and phase (phase is not available for all results) for up to 8 selected harmonics. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 108
V fundamental in this channel. 6.11.2 Comparing Harmonics against Limits 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 108 of 284...
Page 109
208. Harmonics measurements can be exported to a .CSV format textual file on an external 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 110
7. The file is then written. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 110 of 284...
Page 111
NO TRACES SELECTED is displayed centered in the chart. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 112
7. The file is then written. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 112 of 284...
Page 113
You do not have to start RUNNING unless you specifically stopped it. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 114
0.5A value, if higher than 0.5A then it will be higher (by an amount set by the scaling required) and will be lower if below 0.5A. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 115
6.6.6. While measurements are suspended, the historical record still runs, but the traces are blanked during this time. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 116
You may zoom in further by pressing the ZOOM+ again, or may zoom back by pressing the ZOOM- button. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 117
USB drive attached to the front panel USB port. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 118
7. The file is then written. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 118 of 284...
Page 119
LOADZ:CHn:ACDC LOADZ:CHn LOADZ:CHn:H1 MOTOR:WATTS MOTOR:SPEED MOTOR:TORQUE MOTOR:SLIP 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 119 of 284...
Page 120
Note: You do not have to be viewing a SCOPE screen for signals to be captured. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 121
If the period is shown colored orange, the unit is not locked closely enough to allow an accurate representation of the signals. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 122
The button in the OFFSET column allows you to set the signal level which will correspond to the centerline of the chart. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 123
7. The file is then written. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 123 of 284...
Page 124
When changes are made any previously captured traces are cleared. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 125
Pressing the POSITION button allows you to set where the trigger position is located on the unzoomed chart. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 126
Changing traces does not clear any previously made scope capture; it only affects how it is displayed. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 127
The oscilloscope may be run in either SINGLE or CONT modes similarly to a normal oscilloscope. Pressing the CONT button starts continuous oscilloscope captures. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 128
2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 129
Notes: If no signals have been captured then NO DATA CAPTURED is displayed on the chart and the chart is blank. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 130
The screen below shows the capture of an inrush current zoomed in on the initial inrush peak. When the cursor is shown: 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 130 of 284...
Page 131
1. Insert the drive into the front panel USB port and wait for it to be READY. 2. Display the screen to be recorded. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 132
This phase is one period of the user set resolution frequency in length (for 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 133
‘real-time’ to prevent this. These 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 134
The CHANNEL button allows you to select which channel you are viewing the spectrum of. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 135
4. Press the SCREEN IMAGE button. The screen changes to allowing you to enter a file name. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 136
7. The file is then written. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 136 of 284...
Page 137
For a VPA total, this is determined from the polarity of the sum of the channel VAR results. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright...
Page 138
Wye Voltage Delta Voltage Currents 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 138 of 284...
Page 139
VAR for the channels in the VPA 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 140
‘1’ and signal ‘3’ as having a 240° relationship to signal ‘1’, whereas mathematically 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 141
Phase is relative to the A fundamental in the selected channel. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 142
These are not relative measurements. The results are in the units of Amps. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 143
Multiple M2000 Instrument Applications Synchronizing Measurements between VPAs 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 143 of 284...
Page 144
DC offset. Thus, you must ensure that there are no DC signals present on the input terminals of the unit that should not be compensated for. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 145
M2000. This section is 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 146
2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 147
This allows you to inspect the supply voltage to 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright...
Page 148
SINGLE button label changing to WAITING. 5. Apply power to the device. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 149
The example screen below shows the POWER DATA screen which will be used for this purpose. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 150
2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 151
6. The file is then written. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 151 of 284...
Page 152
To manually control starting and stopping of any integration, select MANUAL as the METHOD setting on this screen. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 153
Integration Info Area and press the START button on that screen. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 154
3. Press the DRIVE INFO area of the screen, the screen shows the file export/import screen. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 155
7. The file is then written. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 155 of 284...
Page 156
2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 157
45Hz/20ms setting should be considered. FUND Setting 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 158
The small inaccuracy of the compensation is included in the reported error figure. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 159
1 second interval) and with the desired data being logged (at least the Watts result is recommended). 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 160
START level of 0.0W then effectively the measurement period will be totally manually controlled. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 161
EN50564 5.3.2a method or for the entire measurement 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 162
EN50564:2011 para. 4.4.1. The measurements are maintained in real-time during the 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 163
The file is then written. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 163 of 284...
Page 164
5th harmonic when EN61000-3-12 is configured, are not available. When available, the 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 165
EUT so V@LOAD should be selected. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 166
However you must ensure that the 1.5 second filtering 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 167
With the M2000 there is no need to repeat the entire assessment process to accommodate this. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 168
EUT is a “balanced load”. The definition of 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 169
EUT to within the limits stated in EN61000-3-2 or EN61000-3-2 -12. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 170
Type card in the M2000 is recommended. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 171
Often this is of a substantial effect at these accuracy levels. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 172
EN61000-3-2 clause A.2d, see below for a 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 173
2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 174
3. Configure the channels you wish to use as needed (see section 12.4.2.1). 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 175
EN61000-4-7 (and so to the selected EN61000-3 standard) being available almost immediately. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 176
ASSESS EUT button is pressed when no prior assessment has been performed. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 177
As required by EN61000-3-2 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 178
An example of this screen is shown below. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 179
THC and PWHC assessment measurements as well as affecting the limits applied. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 180
1 second assessment time, then 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 181
2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 182
M2000 does not enforce this and in general the EUT will require longer than this to stabilize. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 183
A.2c throughout the assessment. Note that when performing EN61000-3-12 assessments no checking of the supply voltage harmonics is performed. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 184
5. The actual measured THC (Total Harmonic Current, in Amps) of the EUT. The minimum- maximum THC measured during the assessment. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 185
Notes: If nothing is shown below a FAIL or FAILING status then the channel selected to be displayed is not causing the failure 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 186
EN61000-3-2 harmonics as that data is the POHC of the average harmonics 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 187
ASSESS button, and allows you 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 188
2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 189
Always a blank record 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 190
EN61000-3-12 assessment but record is always present failure code is 0 if no failure or 1 if failed to meet the requirement 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 191
9, “USB Drive Usage” on page 208 of this document. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 192
AC or a DC supply is used by the equipment. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 193
≤150kHz as required in the Airbus requirements because of the 300kHz BANDWIDTH setting used). 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 194
≤150kHz as required in the Airbus requirements because of the 300kHz BANDWIDTH setting used). 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 195
If you wish to record the measurements made for documentation purposes, it is recommended to either: 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 196
2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 197
2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 198
2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 199
COUPLED data, and showing the full DETAILED information. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 200
(for a 1-phase supply there is only a single phase). 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 201
DHCP or FIXED IP LAN modes instead of AUTO-IP. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 202
M2000 on which it is being displayed). Typically this can take a few seconds to respond 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 203
2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 203 of 284...
Page 204
M2000 within the group always includes the data from all members of the group. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 205
LAN interface delay. The synchronizing information passed between multiple M2000’s is 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 206
It should be noted that this error is typically far less 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 207
2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 208
Harmonic Limits file. Importing factory updates. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 208 of 284...
Page 209
USB2.0 specification. The drive should be removed and after 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 210
There are several that are free of charge available on the internet. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 211
This initiates a screen that allows you to export files. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 212
Any voltage and current harmonics limits defined. The measurement preference settings. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 212 of 284...
Page 213
VPA data section (only VPA1 has measurements available in this example). DATA,VPA1,VPA2,VPA3 Frequency,+50.0373E+0,, V13(pk),+0.00000E-9,, 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 213 of 284...
Page 214
Aphase H3,,+99.8311E+0,, Vrms H4,,+52.4739E-3,, Vphase H4,,+170.166E+0,, Arms H4,,+503.284E-6,, 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 214 of 284...
Page 215
32769 records. The first record contains a heading describing each column. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 216
7.00548E-3,-6.89178E-6,, +24.4139E-6,+570.967E-3,+4.02716E-3,+8.12532E+0,+3.69523E- 3,+2.50181E-3,-3.62742E-6,, +29.2967E-6,+691.224E-3,+4.89038E-3,+8.05385E+0,+4.51385E- 3,+7.29706E-3,+531.195E-9,, +34.1795E-6,+819.305E-3,+5.75712E-3,+7.98462E+0,+5.39458E-3,- 7.69836E-3,-2.13687E-6,, 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 216 of 284...
Page 217
This initiates a screen in which you press the IMPORT FILE button to initiate the import file screen. An example of this is shown below. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 218
You can create an ASCII format harmonics limits file for either voltage or current limits on a computer using a plain text editor such as Windows Notepad. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 219
22pix text size: includes the cell to the right of the target cell. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 220
7. A blank field (no characters) 8. A blank field (no characters) 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 221
3. Text which is 28pix high 4. Text which is 36pix high 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 222
This sets the screen to be similar to that shown below. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 223
Data logging is not synchronous to any measurement period, but logs the latest measurements at each respective time interval. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 224
If data is selected to be written to INTERNAL memory then there are no other buttons shown in the FILE area. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 225
The HEADER button allows you to select if the file will start with a header record or not (this is only available for the ASCII FORMAT setting). 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 226
0 is entered then 0.002 seconds is used. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 227
Only measurements which are presently configured to be measured can be selected while in this screen. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 228
The CLEAR button is only shown when a data log has been completed but no data from it has been saved to a drive. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 229
FAT32. In this case, you should export the data in binary format and externally convert it into ASCII using the supplied application. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 230
5Mbytes/sec (1.25Mdata/sec), so if data logging at the maximum rate (0.002sec or 500/sec) 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 231
For best speeds, use a portable hard drive not a flash drive. Hard drives typically have more consistent and faster write speeds. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 232
If possible, use the M2000 USB interface instead. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 233
M2000 is enabled. Generally, this signal is not used by the computer so is not often a cause for concern. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 234
Cable required CAT5 or CAT5e, UTP or STP Cable Length <100m (per standard) 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 234 of 284...
Page 235
M2000 should be able to pass through a router, so the M2000 can be made local to a portion of a network. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 236
>1 minute. In this case, 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright...
Page 237
USB Device Enumerates as a Human Interface Device (HID) of the Vendor-specific type 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 238
To close the connection the following is an example. HidUart_Close(usbdevice) 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 238 of 284...
Page 239
Timer.Stop() Exit Function End If End While 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 239 of 284...
Page 240
Return Value: HID_UART_STATUS = HID_UART_SUCCESS, HID_UART_INVALID_DEVICE_OBJECT, HID_UART_INVALID_PARAMETER, HID_UART_DEVICE_ACCESS_ERROR HidUart_Open 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 240 of 284...
Page 241
2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 242
DWORD writeTimeout) device—Device object pointer as returned by HidUart_Open(). 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 242 of 284...
Page 243
4095 characters in total. There is no action taken or error 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright...
Page 244
Examples of valid KEYWORD fields are: 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 244 of 284...
Page 245
Examples of valid NR3 fields are: 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 246
Except for a STRING command field, a blank field may contain one or more whitespace characters but no other characters. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 247
2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 248
2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 249
5: An interface command field was expected but not found 6: An interface command field was found but not expected 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 250
Responds with the STRING format present time (as selected format). E.g. HH:MM:SSam 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 251
0 when a data log is started. 0: There has been no data logging error 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 252
3: A EN50564 5.3.2d measurement is in progress but is waiting for the start detection level (no data is available). 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 253
2. Send the required measurement configuration commands (in any order) 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 254
0.0, or +1e-3 to +1e6 Otherwise turns on voltage scaling using s 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 255
Full synchronization with VPA2 (not valid if v = VPA1 or VPA2) 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 256
Sets that integration will be automatically timed (i=1) or manually stopped and started (i=0) 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 257
0 to 99 0 to 99 0 to 99 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 257 of 284...
Page 258
Responds with the NR1 efficiency group setting for VPA v: 0: None 1: IN 2: MIDDLE 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 258 of 284...
Page 259
Responds with the NR1 efficiency group setting for motor power: 0: None 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 259 of 284...
Page 260
5: Speed derived from the electrical drive frequency measured in VPA2 (only valid if VPA2 configured for use) 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 261
Responds with the NR3 setting for the torque measurement offset 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 261 of 284...
Page 262
V or A Saves all voltage (i=V) or current (i=A) harmonic limits 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 263
(just=1) or right (just=2) color sets the text color (R:G:B values) 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 264
(responds with 0) or failing (responds with 1) 0 to 500 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 265
EN50564 at any time during the measurement (responds with 1) or not (responds with 0) 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 266
Data is the voltage measured between phases A and C of the selected VPA 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 267
Data is as Hn but expressed as a percentage of the ACDC amplitude 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 268
VOLTS:An:pBC:FF AMPS:CHn:PK-VLY AMPS:An:INTAVG:H1 VOLTS:CHn:INRUSHVLY VOLTS:An:pBC:CF AMPS:CHn:HIPK AMPS:An:CHARGE 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 268 of 284...
Page 269
VAR:An:pAC:ACDC VA:An:pAC:AC PF:An:pAC:INTAVG WATTS:An:pBC:INRUSH VAR:An:pAC VA:An:pAC:ACDC PF:An:pAC:H1 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 269 of 284...
Page 270
STBYERR:CHn:STBYMIN VAR:CHn:INTAVG:H1 VA:An:pX:DC PF:CHn:INTAVG:H1 VAR:CHn:BOUGHT VA:An:pX:AC PF:CHn:H1 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 270 of 284...
Page 271
PK-VLY Data is the difference between peak and valley peak 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 271 of 284...
Page 272
AMPS:An:PK VAR:An:pX:ACDC PF:CHn:ACDC VOLTS:An:pAB:THDsig AMPS:An:H1 VAR:An:pX PF:CHn 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 272 of 284...
Page 273
LOADZ:CHn:H1 AMPS:An:pX WATTS:An:ACDC VA:An:pX:H1 AMPS:An:pX:CF WATTS:An VA:An:pAC:DC 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 273 of 284...
Page 274
M2000 accuracy is negligible at the adjustment levels. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 275
10.25Kohm load for option X or a 102.5Kohm load for option V. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 276
6. (If required) Press the SET DATE and TIME button. This will set the date and time of the M2000to that of the computer. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 277
When the requested input has been provided, press the START ADJUSTMENT STEP button. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 278
Using this will assist you in determining if you are satisfied with the 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 279
WARNING: Do not remove power from the M2000 during or until at least 10 seconds after completing adjustments. 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual...
Page 280
Quality Assurance Inspector Adaptive Power Systems Mark of Compliance 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 280 of 284...
Page 281
Features ................20 Warranty ................11 Weight ................53 2019 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS. Entire Contents Copyright M2000 Series Power Analyzers Operation Manual Page 281 of 284...
Page 284
ADAPTIVE Power Systems Worldwide Supplier of Power Equipment Copyright 2015, Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS.