Medisana IN 500 Compact Manual page 30

Table of Contents

Advertisement

Available languages

Available languages

GB
4 Miscellaneous
4.5
Replace
the fuses
(for art. 54530
IN 550 Pro
only)
4.6
Directives /
norms
4.7
Disposal
24
Inside the compressor unit of IN 550 Pro two fuses (type T1AL, 250V~)
are present to protect the electrical components. Make sure, that the app-
liance is only connected to mains outlets, which are corresponding to the
specifications mentioned in chapter 4.8 Technical specifications. If it is
necessary to replace the fuses, follow the below steps:
1.
Open the mains cable com-
9
partment lid
on the rear
side of the compressor.
2.
By turning the fuse cover
counterclockwise (e.g. with
a screwdriver) and im-
pressing it at the same
time (see alongside mar-
king), you can open the
fuse chambers.
Replace the old fuses by
3.
new ones of the same type
and close the lid again.
This inhalation device meets the requirements of the EU standard 93/42/
EWG (medical device directive) and of the standard EN 13544-1:
2007 Respiratory therapy equipment - Part1: Nebulising systems and
their components.
Electromagnetic compatibility:
The device complies with the EN 60601-1-2 standard for electromagnetic
compatibility. You can find details on this measurement data on the sepe-
rate enclosure.
This product must not be disposed of together with domestic waste. All
users are obliged to hand in all electrical or electronic devices, regardless
of whether or not they contain toxic substances, at a municipal or commer-
cial collection point so that they can be disposed of in an environmentally
acceptable manner. Consult your municipal authority or your dealer for in-
formation about disposal.
g
Rear view of compressor IN 550 Pro
i
E

Advertisement

Table of Contents
loading

This manual is also suitable for:

In 550 pro5452054530

Table of Contents