Internal Lift-Off Protection; Rubber Ring For Mounting - Sartorius PR 6202 Series Installation Manual

Mounting kits for load cell
Table of Contents

Advertisement

Available languages

Available languages

Mounting Kits PR 6002
2.3

Internal Lift-off Protection

All mounting kits are equipped with an internal lift-off protection; i.e. no bore holes other than the
mounting holes for the kit are required.
The Lift-off protection is realized by a bolt (2), nut (1) and a retainer (3). Bolt and nut are delivered in a
bag.
Mount the lift-off protection after installation/alignment of the load cell.
Do not turn the slots by more than 30° from the vertical.
2.4

Rubber Ring for Mounting

1...10 t
The rubber ring for mounting is used for vertical installation of the load cell.
Place the rubber ring on the lower load disk.
Position the load cell between the load disks.
Lower the vessel and tighten the mounting screws.
Pull out the four tabs on the rubber ring by approx. 15 mm so that the rubber ring is removed from
between load disk and load cell.
Then tear or cut off the rubber ring to remove it completely.
EN-4
3
25...50 t
Installation Manual
1
2
Sartorius