How to proceed if a repair is necessary ................41 Dismount..........................42 Dismounting steps......................42 Disposal ......................... 42 Supplement ..........................43 Technical data ........................ 43 Dimensions ........................47 Trademark ........................48 OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...
This arrow indicates a single action. Sequence of actions Numbers set in front indicate successive steps in a procedure. Battery disposal This symbol indicates special information about the disposal of bat- teries and accumulators. OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...
During work on and with the device, the required personal protective equipment must always be worn. Appropriate use OPTISOUND 3030 C is a sensor for continuous level measurement. You can find detailed information about the area of application in chapter "Product description". Operational reliability is ensured only if the instrument is properly used according to the specifications in the operating instructions manual as well as possible supplementary instructions.
Installations in the US shall comply with the relevant requirements of the National Electrical Code (ANSI/NFPA 70). Installations in Canada shall comply with the relevant requirements of the Canadian Electrical Code. OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...
Housing lid with integrated display and adjustment module (optional) Housing with electronics 3 Process fitting with transducer Type label The type label contains the most important data for identification and use of the instrument: • Instrument type • Article and serial number device • Article numbers, documentation OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...
Technical data: For example approvals, process temperature, process fitting/material, signal output, voltage supply, protection Principle of operation Application area OPTISOUND 3030 C is an ultrasonic sensor for continuous level measurement. It is suitable for liquids and solids in practically all industries. Functional principle The transducer of the ultrasonic sensor transmits short ultrasonic pulses to the measured product. These pulses are reflected by...
Page 9
• Relative humidity 20 … 85 % Lifting and carrying With instrument weights of more than 18 kg (39.68 lbs) suitable and approved equipment must be used for lifting and carrying. OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...
Prior to setup you have to replace these protective caps with ap- proved cable glands or close the openings with suitable blind plugs. Measuring range The reference plane for the measuring range is the lower edge of the transducer. OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...
Page 11
(max. measuring distance) Measuring range Pressure/Vacuum Gauge pressure in the vessel does not influence OPTISOUND 3030 C. Low pressure or vacuum does, however, damp the ultrasonic pulses. This influences the measuring result, particularly if the level is very low. With pressures under -0.2 bar (-20 kPa) you should use a different measuring principle, e.g. radar or guided radar (TDR). OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...
Mounting instructions Installation position When mounting the OPTISOUND 3030 C, keep a distance of at least 200 mm (7.874 in) to the vessel wall. If the sensor is installed in the center of dished or round vessel tops, multiple echoes can arise.
Page 13
500 mm 250 mm/10" 600 mm Fig. 7: Deviating socket dimensions Sensor orientation In liquids, direct the sensor as perpendicular as possible to the prod- uct surface to achieve optimum measurement results. OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...
Page 14
Vessel installations such as for example, ladders, limit switches, heat- ing spirals, struts etc. can cause false echoes that interfere with the useful echo. Make sure when planning your measuring site that the ultrasonic signals have a "clear view" to the measured product. OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...
Page 15
If there are agitators in the vessel, a false signal suppression should be carried out with the agitators in motion. This ensures that the interfering reflections from the agitators are saved with the blades in different positions. Fig. 12: Agitators Inflowing medium Do not mount the instruments in or above the filling stream. Make sure that you detect the product surface, not the inflowing product. OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...
Page 16
(TDR). By using a standpipe (surge or bypass tube), the influence of vessel Standpipe measurement installations, foam generation and turbulence is excluded. Standpipes must extend all the way down to the requested min. level, as measurement is only possible within the tube. OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...
Page 17
Fig. 14: Standpipe in the tank Vent hole ø 5 … 10 mm OPTISOUND 3030 C can be used from tube diameters of 100 mm. Avoid large gaps and thick welding joints when connecting the tubes. Generally carry out a false signal suppression.
No grease should be used when screwing the NPT cable gland or steel tube into the threaded insert. Standard grease can contain additives that corrode the connection between threaded insert and housing. This would influence the stability of the connection and the tightness of the housing. OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...
9. Check the hold of the wires in the terminals by lightly pulling on them 10. Connect the screen to the internal ground terminal, connect the external ground terminal to potential equalisation OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...
Electronics and connec- tion compartment Display I 2 C Fig. 17: Electronics and connection compartment, single chamber housing Spring-loaded terminals for voltage supply Ground terminal for connection of the cable screening OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...
Page 22
Plug connector for service interface Connection compartment I 2 C Fig. 21: Connection compartment, double chamber housing Spring-loaded terminals for voltage supply Plug connector for service interface Ground terminal for connection of the cable screening OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...
Fig. 22: Wiring plan - double chamber housing Voltage supply, signal output Switch-on phase Switch-on phase After connecting OPTISOUND 3030 C to power supply or after a voltage recurrence, the instrument carries out a self-check for approx. 30 seconds: •...
The display and adjustment module is powered by the sensor, an ad- ditional connection is not necessary. Fig. 23: Insert display and adjustment module Note: If you intend to retrofit the instrument with a display and adjustment module for continuous measured value indication, a higher lid with an inspection glass is required. OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...
Approx. 60 minutes after the last pressing of a key, an automatic reset to measured value indication is triggered. Any values not confirmed with [OK] will not be saved. Setup steps Address setting HART In HART-Multidrop mode (several sensors on one input) the address multidrop must be set before continuing with the parameter adjustment. You will OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...
Page 26
1. Move from the measured value display to the main menu by pushing [OK]. ▶ Basic adjustment Display Diagnostics Service Info 2. Select the menu item "Basic adjustment" with [->] and confirm with [OK]. Now the menu item "Min. adjustment" is displayed. OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...
Page 27
Enter the requested parameters via the appropriate keys, save your settings and jump to the next menu item with the [->] key. OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...
Page 28
In this menu item you can enter an unambiguous designation for the sor TAG sensor, e.g. the measurement loop name or the tank or product des- ignation. In digital systems and in the documentation of larger plants, a singular designation should be entered for exact identification of individual measuring points. OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...
Page 29
100 % = 100.0 l Display - Backlight A background lighting integrated by default can be adjusted via the adjustment menu. The function depends on the height of the supply voltage. See "Technical data/Voltage supply". OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...
Page 30
Up to 3000 measured values are recorded (depending on the sensor) when starting a "Trend curve". Then the values can be displayed on a time axis. The oldest measured values are always deleted. OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...
Page 31
A false echo memory should be created with low level so that all potential interfering reflec- tions can be detected. False signal suppression Change now? Proceed as follows: OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...
Page 32
Service - Extended set- The menu item "Extended setting" offers the possibility to optimise ting OPTISOUND 3030 C for applications in which the level changes very quickly. To do this, select the function "Quick level change > 1 m/min.". Extended setting quick level change > 1 m/min.
Page 33
If the function "Reset" is carried out, the sensor resets the values of the following menu items to the reset values (see table): Function Reset value Sensor address Max. adjustment 0 m(d) Sensor-specific basic adjustment. OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...
Page 34
3.50: • Deutsch • English • Français • Espanõl Depending on the sensor type, see chapter "Technical data". Special parameters are parameters which are set customer-specifically on the service level with the adjustment software PACTware. OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...
Page 35
The following data are read out or written with this function: • Measured value presentation • Adjustment The 4 … 20 mA signal of the sensor is switched off. The sensor uses a constant current of 4 mA. The measuring signal is transmitted exclusively as a digital HART signal. OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...
Page 36
Serial number: 8-digit number, e.g. 12345678 Instrument type Serial number • Date of manufacture: Date of the factory calibration • Software version: Edition of the sensor software Date of manufacture Software version OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...
Linearisation curve Sensor-TAG Linear Sensor Display Basic adjustment ▶ Display Diagnostics Service Info Displayed value Display unit Scaling Scaled Volume 0 % = 0.0 m³ m³ 100 % = 100.0 m³ OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...
Page 38
Copy sensor data? Enable? Address 0 Info Basic adjustment Display Diagnostics Service ▶ Info Instrument type Date of manufacture Last change using PC Sensor characteristics Software version Display now? Serial number 12345678 OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...
They are thus available for multiple use or service purposes. If OPTISOUND 3030 C is equipped with a display and adjustment module, the most important data can be read out of the sensor into the display and adjustment module.
Exchange the instrument or send it in for repair greater than 22 mA sensor defective or less than 3.6 mA In Ex applications, the regulations for the wiring of intrinsically safe circuits must be observed. OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...
If there is no electronics module available on site, one can be ordered from the Krohne agency serving you. How to proceed if a repair is necessary If it is necessary to repair the instrument, please contact the responsi- ble Krohne agency. OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...
Pass the instrument directly on to a specialised recycling company and do not use the municipal collecting points. If you have no way to dispose of the old instrument properly, please contact us concerning return and disposal. OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...
Ʋ HART value (Primary Value) Distance to the level Ʋ HART value (Secondary Value) Temperature Ʋ HART value (3rd Value) Distance to the level - scaled Glass (with Aluminium and stainless steel precision casting housing) OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...
Page 44
Adjustment time > 3 s (dependent on the parameter adjustment) Incl. non-linearity, hysteresis and non-repeatability. Time to output the correct level (with max. 10 % deviation) after a sudden level change. OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...
Page 45
-20 … +70 °C (-4 … +158 °F) adjustment module Material Ʋ Housing Ʋ Inspection window Polyester foil Relating to the nominal measuring range. Tested according to the guidelines of German Lloyd, GL directive 2. OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...
Page 46
Pollution degree Protection class For instruments with national approvals such as e.g. according to CSA, only available at a later date. A suitable cable is required for maintaining the protection rating. When used with fulfilled housing protection OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...
Fig. 28: OPTISOUND 3030 C Mounting strap Compression flange Dead zone: 0.6 m (2 ft) Measuring range: with liquids up to 15 m (49.21 ft), with solids up to 7 m (22.97 ft) OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...
9 Supplement Trademark All the brands as well as trade and company names used are property of their lawful proprietor/ originator. OPTISOUND 3030 C • Two-wire 4 … 20 mA/HART...