Table of Contents

Advertisement

To set, modify or delete the User-ID, simply place the cursor in the User-ID box and write a new
identifier or clear the text. The new identifier is written to the instrument's memory as soon as return or
enter is pressed, or if the mouse is left-clicked outside of the box.
The User-ID is written to non-volatile memory, so it is retained even in the event of a reset or battery
failure.

8.3.4 Recording Field

Note:
When using the instrument with WiFi™ connectivity, do not use any of the controls in the
Recording field. The instrument will be completely controlled by the server, including starting the
recording, erasing memory if required, and resetting the time.
The recording field has three buttons:
Record
recording. The new record is placed in memory immediately after the last one. The recording
stops when the memory is full, or when the Record button is pressed again. The recording
parameters, such as number and types of values recorded...etc. must be adjusted prior to
starting the recording.
AutoRec
mode. In the Auto-Rec mode the instrument stays semi-dormant, sampling the acceleration or
velocity in the background, as long as it does not detect an acceleration or velocity value
beyond an adjustable threshold. When the signal value goes over the threshold, the instrument
wakes up and starts recording, using the recording parameters in effect in the instrument's
configuration. When the signal value goes and stays below the threshold for a set length of
time,
the
recording
When the Auto-Rec mode is selected, the instrument will first wait for the signal to rise above
the set threshold before starting the first recording. The next period of inactivity will stop the
recording, and so on until the user exits the Auto-Rec mode, or the memory is exhausted. A
new record is created every time the threshold is passed and the instrument is not already
recording.
The threshold is expressed in ��/��
Note:
The threshold T can only be positive. When recording raw signals or signal statistics
the recording is triggered as soon as the signal goes higher than +T, or lower than −T. When
recording RMS levels, the recording is started as soon as the RMS level goes higher than +T
Note:
Timer
automatically. This is only allowed when a recording is not already in progress. To set the
timer, press the Timer button. Then set the recording date and time, and press the OK button
to arm the timer. The timer will start an immediate recording, or start the AutoRec mode,
depending on the settings.
Erase
in the instrument, making room for new records.
The Memory Depth bar provides two indications:
The percentage of memory already used (indicated by the green bar). The bar is completely
grey when the memory is empty and completely green when the memory is full.
User's Manual
The Record button is used to immediately create a new record and start
The Auto-Rec button is used to place the instrument in or out of the Auto-Rec
stops
and
The Timer button is used to set a time and date to start the recording
The Erase button erases unconditionally all the data that has been recorded
the
instrument
goes
for acceleration and in ��/�� for velocity.
2
back
to
sleep.
RMS
.
24

Advertisement

Table of Contents
loading

Table of Contents