Installation and Operating Specification EDR
Pressure connections "+, -"
Housing
Use position
Protection rating
Weight
Permissible ambient temperature
Permissible medium temperature
Permissible temperature range for storage and
transport
Permissible rel. humidity
One-sided permissible overload
Burst prssure
Maximum cross section of terminals
Interference emission
Interference immunity
Accuracy and measuring ranges
Pressure measuring range max.
Tolerance zero point max.
Tolerance full scale max.
Resolution
Total of linearity, hysteresis and repeatability max.
Long term stability according to DIN EN 60770
Temperature coefficient typical
Temperature coefficient max.
Temperature coefficient sensitivity typical
Temperature coefficient sensitivity max.
The accuracy data are percentages and refer to the maximum possible measuring range
of the respective type.
Test conditions: 25 °C, 45 % RH, voltage supply 12 VDC
*) Calibration is possible to reduce the zero point error
L-BAL-E292-GB 1841 Index 001
*)
Hose connectors d = 5 / 6 mm
PC (polycarbonate)
Fire protection classification UL94V0
vertical (measuring depends on position)
IP54 according EN 60529
approx. 230 g
(0.50 lb)
-10...60 °C
(14...140 °F)
-10...70 °C
(14...158 °F)
-30...70 °C
(-22...158 °F)
85 % no condensation
0.1 bar
(80 in.wg)
at room temperature: 0.2 bar
at 70 °C (158 °F): 0.15 bar
2
1.5 mm
/ AWG16
according EN 61000-6-3 (domestic household applica-
tions)
according EN 61000-6-2 (industrial applications)
zero point calibration
Part.-No.
12/15
Enclosure
(0.20 / 0.24 inch)
(80 in.wg)
(60 in.wg)
0...1000 Pa
(0...4.0 in.wg)
%
+/- 0.9
%
+/- 1.3
%
0.1
%
0.6
%
+/- 1.0
% / 10K
+/- 0.2
% / 10K
+/- 0.4
% / 10K
+/- 0.2
% / 10K
+/- 0.4

Advertisement