SCPI Error Messages
−313
Calibration memory lost
Indicates that nonvolatile calibration data used by the *CAL? command has been
lost.
−314
Save/Recall memory lost
Indicates that the nonvolatile data saved by the *SAV? command has been lost.
−315
Configuration memory lost
Indicates that the nonvolatile data saved by the device has been lost. The meaning of
this error is device-specific.
−330
Self-test failed
−350
Queue overflow
A specific code entered into the queue in lieu of the code that caused the error. This
code indicates that there is no room in the queue and an error occurred but was not
recorded.
−360
Communication error
This is the generic communication error for devices that cannot detect the more
specific errors described for errors -361 through -363.
−361
Parity error in program message
Parity bit not correct when data received for example, on a serial port (for example,
a baud rate mismatch).
−362
Framing error in program message
A stop bit was not detected when data was received for example, on a serial port.
−363
Input buffer overrun
Software or hardware input buffer on serial port overflows with data caused by
improper or nonexistent pacing.